Delta OHM HD2102.1 Manual Download Page 13

THE

 

PROBES 

The instrument works with probes of the LP471… series: these are photometric and radiometric 
probes that measure 

illuminance 

(LP471 PHOT), 

irradiance 

(LP471 RAD, LP471 UVA, LP471 

UVB and LP471 UVC), 

PAR

 (LP471 PAR) and 

luminance

 (LP471 LUM 2). All the probes, save 

the LUM 2, are provided with a diffuser for cosine correction.  

Upon turning on 

the instrument automatically detects the probe connected to the input: it is 

sufficient to connect it.  If the instrument is already on, turn it off and back on again in order for the 
probe to be detected. The unit of measurement is determined according to the probe connected to 
the input: In cases where more than one unit of measurement is provided for the same probe, use the 
UNIT key to select the one desired.  
All probes are calibrated in the factory; no calibration is required by the user.  
In addition to instantaneous measurement the instrument calculates the acquired measurement time 
integral Q(t). Some thresholds can be associated with the integrated measurement and with the 
integration time, which can be set in the menu. When exceeded these thresholds cause the 
instrument to stop the integral calculation. 

The probe is detected during turn on, and this cannot be performed when the instrument is 
already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it 
off and on

 
 

Q/T

IME INTEGRATION

 

In addition to instantaneous measurement the instrument calculates the following summation:  

(1)    Q(t) = 

 u(t) t,   

Σ

⋅Δ

Δ

t = 1sec

0

t

 

 

where u(t) is the instantaneous value of the input variable compared to time t. The sampling interval 
is fixed at 1 second. 
As soon as either the Q(t) value or the integration time t reach the set limit, the integration is 
stopped. 
 

Maximum integration time:  100 hours, 00 minutes, 00 seconds

 

Integration interval:        1 second 
 
 

How to set the limits 

To set the integration limit, enter the "

INTG LIMT ###S

" menu item. Use the arrows to set the 

desired value. Confirm by pressing ENTER.  
To set the final integration time, enter the 

TIME LIMT HOUR, TIME LIMT MIN 

and 

TIME 

LIMT SEC 

menu items. Use the arrows to set the desired value in hours, minutes and seconds. 

Confirm each item using ENTER. 
To cancel the limit, set its value to 0. 
 

How to perform an integration measurement 

After setting the limits, as illustrated above, exit the menu and return to normal measurement.  

-      - 

13

Summary of Contents for HD2102.1

Page 1: ...This situation can cause possible differences comparing this manual with the instrument you bought We regret in advance for any possible mistake in this manual Data drawings and descriptions included...

Page 2: ...Photo Radiometer HD2102 1 2...

Page 3: ...en the key is pressed in the menu decreases the current value 9 SERIAL key starts and ends data transfer to the serial communication port 10 MENU CLR Q t key allows access to and exit from the menu If...

Page 4: ...Photo Radiometer HD2102 2 4...

Page 5: ...ey during normal operation starts and ends the saving of the measured or integrated data Q T in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 MENU...

Page 6: ...n the menu When exceeded these thresholds cause the instrument to stop the integral calculation The HD2102 2 instrument is a datalogger It memorizes up to 38 000 samples which can be transferred from...

Page 7: ...detection of the probe connected to the input and sets the instrument ready for normal measurement MIN cm min s 2 3 m m k If during turning on no probes are connected the message PROB ERR is displaye...

Page 8: ...ction cannot be transferred to the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pre...

Page 9: ...previously been set for the calculation by pressing the FUNC ENTER and UNIT Q T keys When used in the menu it clears or cancels the active function REL key In measurement mode it displays the differe...

Page 10: ...e moment of sampling and will turn off immediately afterwards thus increasing the battery life Using the arrows select YES and confirm using ENTER in order to enable the automatic turning off select N...

Page 11: ...d subsequent stop represent a block When the logging function is on the LOG indication is on the display the battery symbol blinks and a beep is issued each time a logging occurs when using an externa...

Page 12: ...ing on the display upon pressing the SERIAL key is printed the variable associated with the probe connected to the instrument with its relevant unit of measurement or the integral Q t Before starting...

Page 13: ...nu When exceeded these thresholds cause the instrument to stop the integral calculation The probe is detected during turn on and this cannot be performed when the instrument is already on therefore if...

Page 14: ...ressing Start Stop Q t in this status to resume the integration press the key again If one or more limits have been enabled in the menu on reaching of the first the calculation stops Pressing FUNC ENT...

Page 15: ...following table lists all error indications and information displayed by the instrument and supplied to the user in different operating situations Display indications Explanation ERR This appears if t...

Page 16: ...LIMT umolm2 integration limit in mol m2 LOG In PROG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO S...

Page 17: ...the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2102 2 is logging and battery voltage falls below the minimum operating level T...

Page 18: ...Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storage conditions...

Page 19: ...USB 2 0 connection does not require the setting of parameters The instruments are provided with a complete set of commands and data queries to be sent via the PC All the commands transferred to the i...

Page 20: ...5 10 3600 seconds WC0 Setting SELF off WC1 Setting SELF on Command characters are exclusively upper case characters Once a correct command is entered the instrument responds with when any wrong combi...

Page 21: ...mand press the MENU key until the _LOG_DUMP_or_ERAS item is displayed and then press on LOG key During data transfer the display shows the message DUMP to stop the data transfer press ESC on the instr...

Page 22: ...g SERIAL again NOTES The print out is formatted across 24 columns Some keys are disabled during serial transmission The following keys work ON OFF HOLD FUNC Max Min Avg and LOG Pressing the HOLD REL a...

Page 23: ...The parameter remains in the memory until replacement of the batteries 4 Launch the DeltaLog9 application and press CONNECT Wait for the connection to occur and follow the indications on the screen F...

Page 24: ...mplete the device will be detected on each new connection automatically In order to check if the entire operation was successful in CONTROL PANEL double click on SYSTEM Select Device Manager and conne...

Page 25: ...ed with instrument off 20 A Mains Output mains adapter 9Vdc 250mA Measuring units lux fcd lux s fcd s W m2 W cm2 J m2 J cm2 mol m2 s mol m2 cd m2 Security of memorized data Unlimited independently of...

Page 26: ...onnector Mains adapter 2 pole connector positive at centre EMC standard regulations Security EN61000 4 2 EN61010 1 level 3 Electrostatic discharge EN61000 4 2 level 3 Electric fast transients EN61000...

Page 27: ...f6 T Temp coefficient 0 05 K Drift after 1 year 1 Working temperature 0 50 C Reference standard CIE no 69 LUMINANCE measurement probe LP 471 LUM 2 complete with SICRAM module and equipped with the in...

Page 28: ...9 99 200 0 1999 9 2000 10000 Resolution mol m2 s 0 01 0 1 1 Spectral range 400nm 700nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1...

Page 29: ...9 Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 400nm 1050nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue...

Page 30: ...solution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 315nm 400nm Peak 360nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fa...

Page 31: ...on W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 280nm 315nm Peak 305nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 2 f4 instrument reading error 1digit f5 fatigue...

Page 32: ...on W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 220nm 280nm Peak 260nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue...

Page 33: ...otted Line Probe Spectral sensitivity LP471 ERY The probe LP 9021 ERY measures the effective total irradiance W m2 eff according to the UV action curve CEI EN 60335 2 27 A particular type of photodiod...

Page 34: ...ngle 2 Measurement range 0 1 cd m2 2000 103 cd m2 LP 471 PAR Quantum radiometric probe for the measurement of the photon flow across the chlorophyll range PAR Photosynthetically Active Radiation 400nm...

Page 35: ...SERIAL INTERFACE AND USB 19 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 21 THE LOGGING FUNCTION ONLY FOR HD2102 2 21 CLEARING THE MEMORY ONLY FOR HD2102 2 21 THE PRINT FUNCTION 22 CONNECTION...

Page 36: ...m d fectueux Dieser Garantieschein muss der Spedition beigelegt werden wenn das Ger t an das Kundendienstzentrum gesandt wird WICHTIG Die Garantie ist nur g ltig wenn dieser Abschnitt bis ins Einzelne...

Reviews: