Expert
Logger
l
Cyclic
Here you can set how often it is sent. The next value is sent after the time
given here has elapsed.
l
Adaptive
This option means that a new value is sent only when the value to be sent has
changed.
l
Never
This option is intended for sending of an initialisation in conjunction with the
next option. The string entered in the
Format
field is then sent only once for
initialisation and never cyclically or adaptively.
l
Initialisation
This option enables you to set initialisation strings, for example, "Meas-
ure:Volt:DC" to switch a multimeter to DC measurement. If you have to send
more than one sequence, enter a number via
Sequence
to determine whether
the string entered in the
Format
field is to be sent first (
Sequence = 1
) or
second (Sequence = 2) and so on. Use the
Time settings
tab to set a pause
after sending an initialisation sequence in order to give the device time to pro-
cess the instruction.
Response expected
Enable this option so that the next output takes place only when the user
has responded to the previous transmission with a string. The response
string itself is not evaluated. Only reception is awaited. If the response con-
tains relevant information, you must define an input channel and then have
the response evaluated.
Time settings tab
Use the
Time settings
tab if you need to use extra pauses for a message,
e.g. to send initialisation to a device that then needs time to perform the ini-
tialisation.
230
Delphin Technology AG
Januar 2018
Expert
Logger
Manual