The pulse generator (
red
curve) is started by the trigger channel
(
blue
curve). When a reset signal (
purple
curve) is received, the
pulse counter is set to "low", but begins immediately to generate new
pulses because the trigger signal remains "High" and the reset signal
is generated only by the edge. The pulse generator stops only once
the trigger signal also goes to "Low". The falling edge of the reset sig-
nal has no effect here as no pulses are being generated at this time.
4.11.8
Integrator
This software channel calculates the area below a curve (
Source
).
Integral
Zeitstempel
Zeitbasis
= Σ
* Δ
*Δ
Vorg nger
aktueller Wert
ä
+
2
Cyclic
: A new value is repeatedly obtained from the source for the given
Period
and the integral is calculated between the current and previous
value. A new value is calculated for each interval and added to the running
total.
Adaptive
: When the source delivers a new value, the integral is calculated
between this value and the previous value. The result is added to the run-
ning total.
Time unit of source
: The calculated values are scaled to the source's time
unit, e.g. in l/min (for litres per minute, enter 60 seconds).
Reset
: When enabled, the calculated integral is output only in the case of a
reset event and then reset internally to zero or the start value.
Refer also to
131