Troubleshooting: Dell PowerVault 124T DLT VS160 Autoloader User's Guide
file:///T|/htdocs/stor-sys/124T/en/124tvs16/trouble.htm[12/5/2012 10:26:08 AM]
If problems recur with the original cartridge or label, replace.
If all previous steps fail, contact
F6
Barcode No Read
Barcode label not present.
Apply valid barcode label to cartridge, or disable the barcode reader
through the Operator Control Panel or Remote Management Unit.
F7
Barcode Mismatch Error
Power cycle the autoloader.
Run the Random Moves test from either the Remote Management Unit
(RMU) or the Operator Control Panel (OCP).
Note: The Random Moves test requires a minimum of two cartridges
present within the autoloader.
If all previous steps fail, contact
F8
Barcode Duplicate Label
Error
Duplicate barcode.
Diagnostic Error Types
F9
Diagnostic Test Failure
Power cycle the autoloader.
Re-run the failing diagnostic test.
If all previous steps fail, contact
FA
Diagnostic Error, Drive
Reporting Write Protect
Verify that valid, not data protected media is present for test.
Run the Random Moves test from either the Remote Management Unit
(RMU) or the Operator Control Panel (OCP).
Note: The Random Moves test requires a minimum of two cartridges
present within the autoloader.
If all previous steps fail, contact
FB
Diagnostic Error, Cleaning
Tape Inserted
FC
Diagnostic Load Tape
Prompt
This is a prompt to manually load a cartridge as required for test.
FD
Diagnostic Unknown Tape
Format
Verify that valid, not data protected media is present for test.
Run the Random Moves test from either the Remote Management Unit
(RMU) or the Operator Control Panel (OCP).
Note: The Random Moves test requires a minimum of two cartridges
present within the autoloader.
If all previous steps fail, contact
FE
Diagnostic Invalid
Configuration
This is an indication that a minimum of two pieces of media are required for test.
Insure at least two cartridges are present in the autoloader and re-run the
failed test.