5.
Manual
Mode
(DDU-240
0 only)
5
Manual Mode
(DDU-2400 only)
The DDU-2400 AED provides a Manual Mode to override the AED features of the defibrillator. Manual Mode provides
an ECG display of patient’s rhythm as well as operator-initiated charge, shock, and disarm functions. Manual Mode
should only be used by qualified medical personnel trained in advanced life support skills and ECG recognition who
want to deliver a shock independent of AED Mode.
WARNING
The DDU-2400 Manual Mode Charge feature is intended
for use only by authorized operators who have been
specifically trained in cardiac rhythm recognition and in
manual defibrillation therapy.
5.1 Entering Manual Mode
When the
Mode Select Icon
is present on the screen, the user may press the
corresponding softkey button to bring up the Mode of Operation screen.
Mode of Operation
AED / Video Mode
Manual Mode
AED / ECG Mode
Manual Mode Softkey
(when in Mode of Operation; DDU-2400 only)
Mode of Operation screen
When Manual Mode is entered, the main screen automatically switches to ECG view and text prompts provide
instruction.
(NOTE: Voice prompts are disabled while in Manual Mode.)
If the pads are not applied, the display
will show a dashed ECG and prompt the user to apply the pads. When the pads are applied, the display shows the
patient’s ECG and heart rate.
Manual Mode
Apply Pads to Patient
Manual Mode
Manual Override
Energy Select
Softkey Icon
Initiate Charge
Softkey Icon
Mode Select
Softkey Icon
Manual Mode: unapplied pads
Manual Mode: pads applied
5.2 Exiting Manual Mode
To exit Manual Mode, press the softkey associated with the
Mode Select Icon
and select a mode of operation.
28