Technical Specifications
DECTRIS PILATUS
®
3 R 100KM
Detector System
Document Version v1.2.2
DECTRIS Ltd.
5405 Baden-DaettwilSwitzerlandwww.dectris.com
Page 1: ...Technical Specifications DECTRIS PILATUS 3 R 100K M Detector System Document Version v1 2 2 DECTRIS Ltd 5405 Baden Daettwil Switzerland www dectris com ...
Page 2: ...Document Version v1 2 2 Copyright 2022 DECTRIS Ltd ...
Page 3: ...NS AND CONNECTORS 9 4 1 PILATUS3 R 100K M Detector 9 4 1 1 Technical Drawing 9 4 1 2 Detector Head 13 4 1 3 Detector Electronics Unit 15 4 1 4 Vacuum Feedthroughs Optional 16 4 2 Detector Control Unit 18 4 2 1 Configuration of the Detector Control Unit 18 4 2 2 Connectors 18 4 2 3 Samba Share 19 4 3 Thermal Stabilization Unit 19 4 3 1 In Vacuum Operation for Detectors with Optional Vacuum Compatib...
Page 4: ...urning Off the Detector 28 7 4 Vacuum Operation for Detectors with Optional Vacuum Compatibility 28 7 5 Storing the Detector 29 7 6 Cleaning and Maintenance 29 8 TROUBLESHOOTING 30 9 CERTIFICATION TESTS 32 10 SERVICE FORM 34 PILATUS3 R 100K M Technical Specifications v1 2 2 ii 34 ...
Page 5: ...Date Changes v1 2 2 2022 02 08 New Server O v1 2 1 2020 09 15 Adjustment of Low Energy Options v1 2 0 2020 07 09 New Linux Distribution v1 1 3 2020 01 16 Low Energy Calibration Option v1 1 2 2019 12 03 New Server J v1 1 1 2019 10 03 New Server O v1 1 0 2019 07 09 Update of corporate design of PILATUS R S and X series v1 0 0 2019 06 28 First Release PILATUS3 R 100K M Technical Specifications v1 2 2...
Page 6: ... use please contact us via telephone e mail or fax 1 2 Explanation of Symbols Danger 0 Danger blocks are used to indicate immediate danger or risk to personnel or equipment Warning 0 Warning blocks are used to indicate danger or risk to personnel or equipment Caution 0 Caution blocks are used to indicate danger or risk to equipment Information 0 Information blocks are used to highlight specific in...
Page 7: ...o the current state of knowledge Damage and warranty claims arising from missing or incorrect data are excluded DECTRIS bears no responsibility or liability for damage of any kind also for indirect or consequential damage resulting from the use of this system DECTRIS is the sole owner of all user rights related to the contents of the manual in particular information images or materials unless othe...
Page 8: ...tor please refer to the order confirmation or check for the presence of the label vacuum tested on the serial number sticker at the detector backside A check in the box of vacuum tested implies the available vacuum option of the PILATUS3 R 100K M detector Please avoid organic materials and highly out gassing compounds inside the vacuum chamber as they tend to deposit on sensitive detector componen...
Page 9: ...nge2 4 5 keV to 36 keV Adjustable threshold range2 2 7 keV to 18 keV Number of thresholds 1 Maximum frame rate 20 Hz Information 1 When using the external trigger or external enable mode the detector will not acquire an image if the ef fective frame rate is above 20 Hz 1 The sensor thickness of your actual system can be found in the order confirmation and in the file header of recorded images 2 Ex...
Page 10: ...thermal stabilization unit 3 1 3 Power Supply Unit The PILATUS3 R 100K M is delivered with the power supply unit TreNew SINPRO MPU130 105 It is a switching power supply Use only the included power supply Please consult the user documentation of the TreNew SINPRO MPU130 105 power supply unit for details 3 1 4 Detector Control Unit The PILATUS3 R 100K M is delivered with the detector control unit DE...
Page 11: ...igger input impedance 50 Ω Detector trigger output 5 V max current 100 mA 3 2 2 Power Supply Unit Table 3 4 Power Supply Unit Ratings Power input 100 VAC to 240 VAC 47 Hz to 63 Hz 1 58 A to 0 64 A Power output 12 V DC max 10 84 A 130 W AC connector IEC 320 C14 input inlet Dimensions 89 5 mm x 49 3 mm x 188 0 mm Weight 0 8 kg 3 2 3 Detector Control Unit Table 3 5 Detector Control Unit Ratings Power...
Page 12: ...wer than 25 during start up use of dry air or nitrogen advised The PILATUS3 R 100K M detector is designed for indoor use only The ambient conditions shown in table 3 7 must be satisfied The stated values are for the ambient conditions Values inside the detector in particular due to the dry air or nitrogen supply are different These are described in section 5 4 and chapter 6 Table 3 7 Detector oper...
Page 13: ...1 Warning 3 Only systems with a vacuum compatibility option are allowed to be operated in vacuum Please contact support dectris com for information regarding vacuum compatibility upgrade For in vacuum operation of the detector following conditions must be fulfilled Table 3 8 In Vacuum Operating Conditions In Vacuum Condition Definition Pressure during operation atmospheric pressure or less than 0 ...
Page 14: ...Scale Format We reserve all rights in connection with this document Without our agreement they may be neither copied nor made accessible to a third party DECTRIS LTD 5400 Baden Switzerland A2 1 1 gek 1 1 0 27 01 2017 prepared for air Zeichnung 162700 M 00 L02 Pilatus3 100K M Air Outline ISO 2768 mK ISO 13920 BF General tolerances Prepared for air views without cover 83 764 487x0 172 33 54 195x0 17...
Page 15: ...O 2768 mK ISO 13920 BF General tolerances Prepared for vacuum All views without cover couplings and mylarframe 83 764 487x0 172 33 54 195x0 172 88 20 4 37 17 to center of sensor 44 to center of sensor G 1 8 10 deep G 1 8 10 deep 42 125 58 5 16 mm to sensor from front edge without cover and mylarframe 11 75 58 View with shifted feet 9 mm threated M6x1 7 threadless 16 138 52 84 to Sensor 52 84 114 6...
Page 16: ...party DECTRIS LTD 5405 Baden Daettwil Switzerland A2 1 2 erb 1 1 0 13 01 2017 Massbild Zeichnung 121400 M 80 L10 Electronics for P3 K Systems ISO 2768 mk ISO 13920 BF General tolerances 50 125 38 5 M6x1 9 5 TIEF 206 6 202 151 155 5 102 95 56 51 3 16 Center of gravity Center of gravity Figure 4 3 Drawing of the PILATUS3 R 100K M Detector electronics unit also printed separately in the user document...
Page 17: ...heit Rougness europ Projektion Bezeichnung Zeichnungsnummer Title Drawing No Datei File Format 17 06 2013 Revision Page Gewicht Pages 1 of 25pin 1of 50pin SUB D F T weight JB 1 210S D25 D50 ISO100 DEC Standard Toleranz F Products SEC02 21 218 F Products SEC02 21 218 All informations provided in this drawing is considered to be confidential proprietary and secret information It is intended only for...
Page 18: ...igure 4 6 The PILATUS3 R 100K M detector head with the protective cover in place front view Danger 1 Danger of electric shock Do not touch the Mylar foil The sensors behind the Mylar foil are operated at high voltages Touching the Mylar foil can cause an electrical shock and damage of the sensors Caution 5 The cover may not protect the detector from a direct beam Figure 4 7 The PILATUS3 R 100K M d...
Page 19: ... protection Be very careful when mounting and removing the Mylar window and avoid touching the sensor under all circumstances as this will most likely result in a damaged sensor Make sure the Mylar window does not tilt inward through the opening at any time Stay clear of the detector volume with any kind of object screw driver finger This operation should be restricted to trained personnel Damage ...
Page 20: ... 100K M detector electronics unit viewed from the front left and back right Table 4 2 Connectors for cables on the front plane of the detector electronics unit Connectors Description DATA1 Data cable to the vacuum feedthrough detector head blue Samtec SQCD ribbon cable POWER1 Power cable to the vacuum feedthrough detector head grey Sub D cable Table 4 3 Status LEDs on the detector back plane LED D...
Page 21: ...se a Lemo Type 00 NIM CAMAC cable Functional ground Information 4 Although the detector might be already grounded via the mounting bolts the detector should be grounded additionally via the functional ground connector at the back to establish a defined grounding FUSE 4 AT Fuse see table 3 3 4 1 4 Vacuum Feedthroughs Optional For vacuum applications separate electrical and cooling feedthrough sets ...
Page 22: ...ional PILATUS3 R 100K M electrical vacuum feedthrough viewed from the vacuum left and air side right Figure 4 12 Picture of the optional PILATUS3 R 100K M cooling vacuum feedthrough set with tubes through a standard DN40ISO KF flange and straight and elbow connectors for the detector head PILATUS3 R 100K M Technical Specifications v1 2 2 17 34 ...
Page 23: ...e interfaces described in section 4 2 2 The detector control unit is optimised for performance and stability of operation In order to achieve these goals we deliver the detector control unit with fixed firmware bios etc and software OS version Figure 4 13 PILATUS3 R 100K M detector control unit as seen from the front Caution 11 Pushing the power button on the front panel longer than 2 seconds will...
Page 24: ...U The Samba service is enabled by default and running after the system boots The storage directory for the images home det p2_ det images which is a symbolic link to home det images can be accessed from a Samba client You need the same user name and password as for the normal console or ssh login From a Linux or Mac OSX terminal window issue the following command smbclient server ip or hostname im...
Page 25: ...on of thermal stabilization unit Use opaque hoses to avoid the growth of algae Do not set the temperature of the thermal stabilization unit below the recommended operating temperature Condens ing moisture can develop and damage the detector 4 3 1 In Vacuum Operation for Detectors with Optional Vacuum Compatibility For in vacuum operation unscrew the self sealing valves from the detector and use va...
Page 26: ...hich are described below 5 2 1 Mounting the Detector Head Figure 5 1 Picture of the PILATUS3 R 100K M detector head with extended mounting brackets viewed from below left and front right Mounting the Detector Head from Above Use the mounting brackets as depicted in figure 5 2 These mounting brackets have to be mounted on the base plate of the detector The detector should be mounted using both oute...
Page 27: ...ounted using the two internal M6x1 threads as shown in figure 5 3 indicated with yellow circles Figure 5 3 Drawing of the PILATUS3 R 100K M detector head base plate a printed copy is supplied in the user documentation folder Left View from below with retracted mounting brackets for mounting from below Right Cross section view through the internal thread to show the 7 mm thread less and the 9 mm th...
Page 28: ...etector control unit and the power supply of the detector have to be connected to a grounded power outlet Although the detector might be already grounded via the mounting bolts the detector should be grounded additionally via the functional ground connector at the back in order to establish a defined grounding 5 4 Connection to Dry Air or Nitrogen Caution 15 Humidity might damage the detector Make...
Page 29: ...r flow is necessary Warning 6 When venting with nitrogen take proper precaution against the risk of asphyxiation caused by oxygen displacement from nitrogen Ensure sufficient ventilation and oxygen level monitoring Use compressed dry air for venting large vessels or in confined spaces 5 5 Connection to Thermal Stabilization Unit The PILATUS3 R 100K M detector is water cooled and must be connected ...
Page 30: ...onal PILATUS3 R 100K M cooling vacuum feedthrough set section 4 1 4 5 6 Mounting the Detector Control Unit Caution 16 Make sure that the detector control unit has adequate ventilation The detector control unit can be mounted in a standard 19 inch rack which has to be properly grounded PILATUS3 R 100K M Technical Specifications v1 2 2 25 34 ...
Page 31: ...tly please refer to chapter 7 and execute the correct start up procedure If the temperature or humidity is out of range the following error message appears _ Camserver Output ERROR temperature too high 31 1C channel 2 Bad return from dcbe_initialize Camera initialization error press enter to exit Make sure that the cooling unit is running at the specified temperature and that nitrogen or dry air f...
Page 32: ...d and mouse to the detector control unit 7 2 Start up Procedure Please use the following start up procedure Turn on the dry air or nitrogen at least 30 min before turning on the detector Then turn ON the power switch at the back of the detector Turn on the detector control unit Wait at least till the detector control unit is booted about 5 min before trying to connect Start a shell The default pat...
Page 33: ...vacuum chamber Connect the detector control unit and the detector data cables Make sure the power switch on the back of the detector electronics unit is OFF Close the vacuum chamber and start to pump down Once the pressure inside the vacuum chamber is below 1 10 2 mbar set the temperature on the thermal stabilization unit to 10 C and turn on the thermal stabilization unit pumping down a warm detec...
Page 34: ...or Please follow these instructions Put the detector in a plastic bag add at least 200 g of drying agent i e silica gel into the bag and seal it air tight Check the humidity and change the drying agent frequently for compliance with the storage requirements in section 3 3 7 6 Cleaning and Maintenance Caution 18 The Mylar foil must not be touched or cleaned If it becomes dirty or is damaged please ...
Page 35: ...he back and on the front panel of the de tector control unit which have to be in the correct position Communication error the detector is not found at start up Data cable is not connected or defec tive The configuration of the Ethernet adapter is wrong Check the connection between detector control unit and detector Make sure that there is a direct peer to peer connection between the detector contr...
Page 36: ...The power cord is not connected or the plug is incompletely inserted The fuse is blown The temperature is over the criti cal limit The thermal protection was triggered Connect the power cord firmly Check the green POWER LED Replace the fuse see table 3 3 Check the thermal stabilization unit The detector will power on again as soon as the temperature is within the allowed operating conditions Image...
Page 37: ...rtification Tests for PILATUS3 R 100K M EN 61000 6 2 2016 IEC 61000 6 2 2005 ed2 0 industrial EN 61000 6 4 2007 A1 2011 IEC 61000 6 4 2011 ed2 1 industrial EN 61010 1 2010 IEC 61010 1 2010 AMD1 2016 EN 61326 1 2013 IEC 61326 1 2012 ed2 0 PILATUS3 R 100K M Technical Specifications v1 2 2 32 34 ...
Page 38: ...S ELA DECTRIS INSTANT RETRIGGER DECTRIS MYTHEN DECTRIS PILATUS DECTRIS QUADRO DECTRIS SANTIS DECTRIS SÄNTIS DECTRIS SINGLA DECTRIS TITLIS DETECTING THE FUTURE 9 2022 DECTRIS Ltd all rights reserved Subject to technical modifications PILATUS3 R 100K M Technical Specifications v1 2 2 33 34 ...
Page 39: ...10 SERVICE FORM PILATUS3 R 100K M Technical Specifications v1 2 2 34 34 ...