Chapter 4 - Technical Specification
4.3.2 Guidance and manufacturer declaration – electromagnetic immunity
TABLE 13: List of electromagnetic immunity compliance
Immunity test
IEC60601 compliance
test level
Electromagnetic environment-guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power supply
lines
±1 kV for input / output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC61000-4-5
±1 kV line to line
±2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
lines
IEC61000-4-11
<5% U
T
(>95%dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
device requires continued operation during mains
interruptions, it is recommended that the device be
powered from an uninterruptible power supply.
<5% U
T
(>95% dip in U
T
)
for 5 seconds
If the user of the device requires continued operation
during mains interruptions, it is recommended that the
device be powered from an uninterruptible power
supply.
Power frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to the application of the test level.
Conductive RF
IEC61000-4-6
3Vrms
150kHz to 80MHz
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2√P
Page 52 of 60
English
Part
Number:
G102748/11_ENG
Copyright
2021
,
Optos
plc.
All
rights
reserved.
Summary of Contents for P200T
Page 1: ...optos com User Guide P200T Optos Panoramic Ophthalmoscope G102748 11_ENG...
Page 4: ...Page 4 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 10: ...Page 10 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 26: ...Page 26 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 58: ...Page 58 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 60: ...G102748 11_ENG English...