Chapter 2 - Introduction
No. Name of Part
Description
5
Image server
Runs the software which transfers patient images from the scan head to the
review client.
6
Customer network
Switches, Firewalls, Internet access, Review clients, Virtual image server (if
required).
7
Patient
environment
The image server, switch, browser-based review clients and any Viewing PCs or
monitors must be positioned outside the patient environment; more than 1.5
meters (5 ft) from the scan head.
Note
The device also includes a user guide, dust cap, power cable and ethernet cable not shown on this
diagram.
A table may also be supplied.
2.6.2 Network specification and characteristics
The following network specification is recommended when connecting the image server to a Wide Area
Network (WAN).
TABLE 1: Network specification
Network specification details
Installation
Scan head and image server PC
Connection type
LAN TCP/IP ipV4
Connection speed
1 Gbps recommended
Connection security
All transport HTTPS by default
Connection ports (defaults)
443, 56620
Unused ports
Locked on scan head and image server PC
Page 23 of 60
Part
Number:
G102748/11_ENG
Copyright
2021
,
Optos
plc.
All
rights
reserved.
English
Summary of Contents for P200T
Page 1: ...optos com User Guide P200T Optos Panoramic Ophthalmoscope G102748 11_ENG...
Page 4: ...Page 4 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 10: ...Page 10 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 26: ...Page 26 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 58: ...Page 58 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 60: ...G102748 11_ENG English...