background image

JUL-2021

MBRTU-TBD is an advanced digital turbidity sensor for water quality monitoring, adopt the principle of scattered light,

the design method of using infrared LED light source and optical fiber conduction light path. The filter design is added

inside, which has strong anti-interference ability. Built in temperature sensor, automatic temperature compensation,

suitable for online long-term monitoring of the environment.

1.  Digital sensor, direct output RS-485 digital signal, support Modbus / RTU

2.  Principle of 90° Angle Scattering Light, the built-in temperature can be compensated automatically;

3.  Optical fiber structure, strong resistance to external light interference

4.  Infrared LED light source, add filter design, anti light interference, good stability

5.  The surface shall be treated with anti-corrosion and passivation

6.  Low power consumption and anti-interference design of internal circuit

Item

Specifications

Output

RS-485,MODBUS/RTU

Measuring method

90° scattered light method

USER GUIDE FOR TURBIDITY
SENSOR WITH MODBUS RTU
OUTPUT MBRTU-TBD

This document is applied for the following products

1. Introduction

2. Specification

Features

Technical Data

Summary of Contents for MBRTU-TBD

Page 1: ...pport Modbus RTU 2 Principle of 90 Angle Scattering Light the built in temperature can be compensated automatically 3 Optical fiber structure strong resistance to external light interference 4 Infrare...

Page 2: ...wer Supply 12 24VDC 10 10mA Size Diameter 30mm Length 166 5mm Protection level IP68 The water depth is 20 meters Service life 3 years or above Cable length 5m Sensor housing material PVC Please wiring...

Page 3: ...probe Suggestion the protective cover of our company should be selected to prevent the influence of microbial attachment on the measurement results Problem Possible Causes Solution The operation inte...

Page 4: ...te the data as the actual value of the standard solution 10 1 2 bytes unsigned short 0x2000 Sensor address R W The default is 1 and the data range is 1 127 1 2 bytes unsigned short 0x2003 Baud rate se...

Page 5: ...4 to m register data Byte m 1 M 2 CRC16 check sum from byte 1 to M When the slave receives an error the slave returns the error 1 2 3 4 5 ADR 83H Information code CRC low byte CRC high byte The first...

Page 6: ...42004 Number of registers 1 Function code 0x06 Default value 1 9600bps Supported values 0 2 4800 19200bps The baud rate can be changed by the upper computer setting and it can work without restart aft...

Page 7: ...nsigned integer data temperature value integer 10 1 decimal place reserved Register returns hexadecimal unsigned integer data turbidity value integer 10 1 decimal place reserved d Humidity measurement...

Page 8: ...d 01 06 10 01 00 00 DC CA Respond 01 06 10 01 00 00 DC CA c Turbidity slope calibration Address 0x1003 41004 Number of registers 1 Function code 0x06 Calibration example calibration in 50NTU turbidity...

Page 9: ...Nam Hung Vuong Res An Lac Ward Binh Tan Dist Ho Chi Minh City Vietnam Tel 84 28 6268 2523 4 ext 122 Email info daviteq com www daviteq com Revision 31 Created Tue Jul 20 2021 3 06 AM by Tien Nguyen Up...