-16-

Circular Leveling Vial

If the circular vial is centered after leveling 

the plate vial, then no further adjustment 

is necessary. If not, proceed with the 

following adjustment.
A.  Using the adjusting pin, provided in 

the case, turn the adjusting screws 

until the bubble is centered in the vial.

Do not over tighten the adjusting screws

B.  Loosen one screw by turning it 1/4 

turn, then tighten the other screw by 

1/4 turn.

Optical Plummet

This adjustment is required to make 

the line of sight of the optical plummet 

coincide with the vertical axis.
A.  Point the optical plummet at a ground 

target. This can be done by adjusting 

the level screws or by loosening 

the instrument fastening screw and 

moving the instrument over the 

target.

B.  Turn the instrument 180° and 

re-sight the target through the optical 

plummet. If the target is still centered, 

no adjustment is necessary. If not, go 

to next step.

D.  Using the adjusting pin, provided in 

the case, turn the reticle adjusting 

screws until half the displacement has 

been removed. Loosen one screw 

1/4 turn, then tighten the opposing 

screw 1/4 turn.

Ground

Target

Stadia

Hairs

half the 

distance

A

A

A

B

L

E.  Repeat steps A through D until there 

is no displacement of the target when 

the instrument is turned about its 

vertical axis.

Collimation Error

•   Attach the instrument on a tripod and 

precisely level.

•   Aim at point A in the distance with the 

normal, direct position of telescope. 

Record the reading of the horizontal 

angle - 

HR-DIRECT

•   Using the reverse position of the 

telescope, take the reading of the 

horizontal angle - 

HR-REVERSE

, then:

 Collimation 

Error C = (HR-Direct -— 

HR-Reverse ± 180°)/2

If 

C <10”

, no adjustment is required. 

If 

C >10”

, the following adjustment is 

required:
•   Adjust the horizontal fine motion in 

the reverse position of the telescope 

so that the reverse reading 

HR-Rev = 

HR-Rev + C

.

•   Remove the protective cover of the 

reticule of the telescope and adjust 

both the left and right adjusting 

screws so that the vertical hair of the 

reticule coincides with object A.

•   Repeat the steps until acceptable 

condition is reached.

Summary of Contents for DT05LS

Page 1: ...DT8 05LS DT8 05 Series Owner s Guide www davidwhite com HOLD R L OSET V ON OFF FOR CUSTOMER SERVICE PARTS AND REPAIR CALL 765 581 4097 IMPORTANT IMPORTANT IMPORTANTE Read Before Using Lire avant usage Leer antes de usar ...

Page 2: ...xtremely rugged field use Like all precision instruments however they should be treated with reasonable care to prolong life and accuracy All instruments are adjusted when they are shipped from the factory It is the customer s responsibility to check and to ensure instruments are adjusted prior to using David White is not responsible for errors caused by instruments that are out of adjustment Cont...

Page 3: ... 3 HOLD R L OSET V ON OFF 2 3 12 13 7 10 9 16 15 14 17 18 4 6 8 7 5 11 1 ...

Page 4: ... and is provided with protection against dust and moisture Once dust or moisture enters into the instrument they will cause damage to the instrument Therefore after using in a humid environment the instrument must be dried immediately and stored in the instrument case Avoid any impact to or dropping of the instrument After severe exterior effects to the instrument it is recommended to carry out an...

Page 5: ... Minimum focus 53 in 1 3 m 53 in 1 3 m 53 in 1 3 m Stadia constant 100 100 100 Electronic Angle Measurement Angle measurement Absolute Encoding Absolute Encoding Absolute Encoding Minimum reading 1 5 1 5 1 5 Direction method H Both sides V Single side H Both sides V Single side H Both sides V Single side Accuracy 5 5 5 Unit of angle Deg mil gon V Deg mil gon V Deg mil gon V Display LCD both sides ...

Page 6: ... Vial Sensitivity Plate vial 30 per 2 mm 30 per 2 mm 30 per 2 mm Circular vial 8 per 2 mm 8 per 2 mm 8 per 2 mm Display Operating Range Operating voltage typical 4 8 VDC 4 8 VDC 4 8 VDC Batteries 4 x 1 5V AA Rechargeable Ni MH 4 x 1 5V AA Rechargeable Ni MH 4 x 1 5V AA Rechargeable Ni MH Operating Temperature range 4 F to 122 F 20 C to 50 C 4 F to 122 F 20 C to 50 C 4 F to 122 F 20 C to 50 C Water...

Page 7: ...ng into Compensator Setting 3 Function button for input while in Instrument Setup HOLD R L OSET V ON OFF HOLD R L OSET V ON OFF HOLD button for reading of horizontal circle for setting or releasing the current horizontal circle value 1 Menu selection button in Instrument Setup 2 Function button for entering while in Instrument Setup 3 Function button for entering into Index Error Setting R L V ON ...

Page 8: ...tery charger The batteries are supplied partially charged To ensure full capacity of the batteries completely charge the batteries before the first use Connect the charger to the AC power source The green light on the charger will luminate Connect the charger to the charging port of the rechargeable battery pack The green light of the charger will turn red indicating the charging process has start...

Page 9: ... Over a Point WARNING Never use force on any parts of the instrument All moving parts will turn freely and easily by hand The DT8 05P and DT8 05LS is equipped with an optical plummet The DT8 05LP is equipped with a laser plummet This eliminates the need to use a plumb bob The instrument is set up over a reference point by looking through an eyepiece and optically lining up over the point DT8 05P a...

Page 10: ...crews Only a firm contact between the screws and the base is necessary If the instrument shifts on the base turn down the screw more firmly by hand Leveling the instrument so the vial bubble remains centered through a 360 rotation of the telescope is the most important operation in preparing to use your instrument When leveling your instrument be sure not to touch the tripod Follow these instructi...

Page 11: ... OSET V ON OFF R L V ON OFF button when the full character display appears and release V HR TILT HOLD R L OSET V ON OFF HOLD R OSET V HOLD R L OSET V and V HR TILT HOLD R OSET V ON HOLD OSET HOLD R L OSET V ON OFF buttons when four beeps are heard The instrument enters into its initial setup mode and the LCD displays SET ZEN VER SET 30 O AUT SET D SET TIL SET NO SET 360 UNIT A TILT SET 0 150 36 10...

Page 12: ...e LCD displays V HR TILT HOLD R L OSET V ON OFF SET ZEN 90 VERTICAL SET 30 OFF AUTO OFF SET DSP 1 SET TILT ON SET NO BEEP TILT 90 00 10 359 59 49 V HR TILT TILT 359 59 49 V HR TILT 65 59 35 359 59 49 V HR TILT 44 537 359 59 49 V HR TILT 92 26 16 V SET 360 UNIT A TILT SET 0 150 36 10 V HR Move the telescope up and down when the instrument is at the normal position The beeper beeps and the LCD displ...

Page 13: ...0 00 02 45 26 10 V HR TILT SET 0 356 65 55 V HR Press HOLD R L OSET V ON OFF HOLD R L OSET V ON OFF R L V ON OFF button again the vertical angle display is resumed 90 TICAL FF O OFF P 1 ON TILT 90 00 10 359 59 49 V HR TILT TILT 359 59 49 V HR TILT 65 59 35 359 59 49 V HR TILT 44 537 V TILT 90 00 02 3 03 05 V HR SET 0 FACE 1 V 96 28 48 FACE 1 V 272 36 06 FACE 2 V 90 00 02 V TILT 90 00 02 356 65 55 ...

Page 14: ...D R L OSET V ON OFF LD R L ET V ON OFF button OFF will be displayed at the position of vertical angle display after a beep release ON OFF HOLD R L OSET V ON OFF OLD R L SET V ON OFF button the instrument is turned off 59 35 59 49 537 59 49 26 16 59 49 08 16 59 49 00 02 59 49 00 02 00 00 272 36 06 FACE 2 V 90 00 02 150 36 10 V 90 00 02 356 65 55 V HL TILT 90 00 02 0 00 00 V HR TILT 90 00 02 45 26 1...

Page 15: ...to ensure the adjustments are correct Order of adjustment 1 Check and adjust the tubular plate vial 17 2 Check and adjust the circular leveling vial 8 3 Check and adjust the optical plummet 11 DT8 05P DT8 05LS Always recheck your adjustments We recommend that any adjustments other than those above be done by a qualified repair technician Tubular plate vial A Place the plate vial parallel to the le...

Page 16: ...the case turn the reticle adjusting screws until half the displacement has been removed Loosen one screw 1 4 turn then tighten the opposing screw 1 4 turn Ground Target Stadia Hairs half the distance A A A A A B L E Repeat steps A through D until there is no displacement of the target when the instrument is turned about its vertical axis Collimation Error Attach the instrument on a tripod and prec...

Page 17: ...he telescope in the normal position at object P and press HR ON OF HOLD OSET HOLD R L OSET V ON OFF to confirm V HR TILT HOLD R L OSET V ON OFF SET ZEN 90 VERTICAL SET 30 OFF AUTO OFF SET DSP 1 SET TILT ON SET NO BEEP TILT 90 00 10 359 59 49 V HR TILT TILT 359 59 49 V HR TILT 65 59 35 359 59 49 V HR TILT 44 537 359 59 49 V HR TILT 92 26 16 359 59 49 V HR TILT 80 08 16 359 59 49 V HR TILT 90 00 02 ...

Page 18: ...thorized after sales service center for Dave White s SitePro instruments In all correspondence and spare parts orders please always include the model number and serial number of the instruments All precision instruments should be cleaned lubricated checked and adjusted ONLY at a qualified instrument repair station or by the manufacturer at least once a year In case of repairs send in the instrumen...

Page 19: ...A 90 DAY LIMITED WARRANTY To make a claim under this Limited Warranty you must return the complete product transportation prepaid For details to make a claim under this Limited Warranty please visit www davidwhite com or call 1 855 354 9881 ANY IMPLIED WARRANTIES SHALL BE LIMITED IN DURATION TO ONE YEAR FROM DATE OF PURCHASE SOME STATES IN THE U S AND SOME CANADIAN PROVINCES DO NOT ALLOW LIMITATIO...

Page 20: ...42346D8895 09 15 Printed in China Dave White s SitePro 7619 S 1150 E Otterbein IN 47970 Tel 1 765 581 4097 ...

Reviews: