Introduction 

 

This Quick Start Guide will provide basic information about 
Datrend Systems’ ES601-CMMS Automated Electrical Safety 
Analyzer.  Please refer to the Operating Manual on the USB that 
accompanied this device for more complete instructions.  The 
steps in this guide will refer you to the appropriate section of the 
Operating Manual, as applicable.  For instructions on installing 
the ES601pc-US program see the 

ES601pc-US Software for 

CMMS systems

 section on the second page. 

 

Unpacking and Inspection 

 

Follow standard receiving practices upon receipt of the 
instrument. Check the shipping carton for damage. If damage is 
found, stop unpacking the instrument. Notify the freight carrier 
and ask for an agent to be present while the instrument is 
unpacked. There are no special unpacking instructions, but be 
careful not to damage the instrument when unpacking it. Inspect 
the instrument for physical damage such as bent or broken 
parts, dents, or scratches, and ensure all accessories are 
present.  If any accessories are missing, immediately contact 
Datrend Customer Service and provide the ES601-CMMS serial 
number. 

 
Starting Up
 (Setup) 

 

Before starting, attach the Power Cord to ES601-CMMS and 
plug in to a power receptacle.  Turn the Power Switch to the ON 
and wait until the MAIN MENU is displayed.  Press System 
Settings to go to the Setup Menu to set defaults. 

  

STARTUP MODE:  
Determines how ES601-CMMS performs after power is 
switched on. Press the Startup Mode button to toggle between: 
• Do not test line voltage at power-on (default) 
• Test line voltage at power-on 

 

PRE-TEST INSPECTION: 
Press to enable/disable prompts for equipment physical 
inspection each time an automated test is started. Default is 
‘enable’. 
 
POST-TEST COMMENT: 
Press to enable/disable prompt for user comments at the end of 
an automated test. Default is ‘enable’. 
 
PROMPT FOR USER TIME: 
Press to enable/disable prompt for user input of labor time at the 
end of an automated test. Default is ‘enable’. 
 
Press More Settings to view settings for MANUAL MODE 
TESTING.  
 
PAUSE AFTER MEASUREMENT:
  
In MANUAL MODE, sets the time delay between measurements 
of a leakage current test. Default 1 second. 
 

 
POWER-UP DELAY:  
In MANUAL MODE, sets the time delay after powering the Test 
Receptacle, before leakage current is measured. Default 2 
seconds. 
 
STOP AFTER SUPPLY ON:
  
STOP BEFORE SUPPLY OFF:  
In MANUAL MODE, determines whether a leakage test 
sequence stops to allow boot-up or shut down of equipment 
under test. To test leakage without stopping for power-
on/power-off of equipment, set to defaults: ‘NO’ and ‘NO’ 
respectively. Refer to operator’s manual for more details. 
 
TEST MULTIPLE RESISTANCES:
 
TEST GROUND WIRE RESISTANCE: 
TEST POINT-TO-PT RESISTANCE:
 
Configures ground resistance test in MANUAL MODE. For 
measurement of power cord resistance, set to defaults: ‘NO’, 
‘YES’, and ‘NO’ respectively. Refer to operator’s manual for 
more details. 
 
Press More Settings to view time/date and other settings.  
 
SET CLK:  
Allows setting the internal real time clock.  Press Set Clk to set 
the current time and date for the device. 
 
FORMAT:  
Select date format for the internal real-time clock.  Press Format 
to toggle between:  
• USA (MMM DD YYYY) (default) 
• Europe (DD MMM YYYY) 
• ISO (YYYY MMM DD) 
 
EDIT ID:  
ES601-CMMS I.D. is set at time of manufacture to "MY ES601”, 
and can be edited using this utility which allows you to assign a 
unique I.D. to identify ES601-CMMS within your equipment 
control system.  Pressing Edit ID will activate the QWERTY 
touch keypad of ES601-CMMS, allowing the user to type in an 
I.D. Press Enter on the keypad to save the unique name and 
return to the Systems Settings Menu. 
 
NETWORK SETUP:
  
Allows you to input Ethernet settings (IP address, netmask, and 
gateway) for ES601-CMMS. Addresses are entered with 
decimal delimiters. A touch keypad that provides for numerical 
input only is displayed in response to pressing any of the 
Change keys.  
 
LABEL PRINTER SETUP:  
Enables and configures output to PM Tag printer (Dymo 
LabelWriter model SE300). Enter title for PM Tags and 
equipment PM interval via this menu. Press Back to save 
changes to PM Tag printer setup. 
 

Quick Start Guide 

Package Contents

 

8001-125 

ES601-CMMS 

3000-009 

POWER CORD 

3140-403 

CABLE, USB A-B MALE, 6' 

3140-475 

KELVIN CLAMP CABLE 

6950-021 

DATREND UNIVERSAL PRODUCT USB DRIVE

 

7006-052 

ACCESSORY BAG 

 
 