Introduction
This Quick Start Guide will provide basic information about
Datrend Systems’ ES601-CMMS Automated Electrical Safety
Analyzer. Please refer to the Operating Manual on the USB that
accompanied this device for more complete instructions. The
steps in this guide will refer you to the appropriate section of the
Operating Manual, as applicable. For instructions on installing
the ES601pc-US program see the
ES601pc-US Software for
CMMS systems
section on the second page.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the
instrument. Check the shipping carton for damage. If damage is
found, stop unpacking the instrument. Notify the freight carrier
and ask for an agent to be present while the instrument is
unpacked. There are no special unpacking instructions, but be
careful not to damage the instrument when unpacking it. Inspect
the instrument for physical damage such as bent or broken
parts, dents, or scratches, and ensure all accessories are
present. If any accessories are missing, immediately contact
Datrend Customer Service and provide the ES601-CMMS serial
number.
Starting Up (Setup)
Before starting, attach the Power Cord to ES601-CMMS and
plug in to a power receptacle. Turn the Power Switch to the ON
and wait until the MAIN MENU is displayed. Press System
Settings to go to the Setup Menu to set defaults.
STARTUP MODE:
Determines how ES601-CMMS performs after power is
switched on. Press the Startup Mode button to toggle between:
• Do not test line voltage at power-on (default)
• Test line voltage at power-on
PRE-TEST INSPECTION:
Press to enable/disable prompts for equipment physical
inspection each time an automated test is started. Default is
‘enable’.
POST-TEST COMMENT:
Press to enable/disable prompt for user comments at the end of
an automated test. Default is ‘enable’.
PROMPT FOR USER TIME:
Press to enable/disable prompt for user input of labor time at the
end of an automated test. Default is ‘enable’.
Press More Settings to view settings for MANUAL MODE
TESTING.
PAUSE AFTER MEASUREMENT:
In MANUAL MODE, sets the time delay between measurements
of a leakage current test. Default 1 second.
POWER-UP DELAY:
In MANUAL MODE, sets the time delay after powering the Test
Receptacle, before leakage current is measured. Default 2
seconds.
STOP AFTER SUPPLY ON:
STOP BEFORE SUPPLY OFF:
In MANUAL MODE, determines whether a leakage test
sequence stops to allow boot-up or shut down of equipment
under test. To test leakage without stopping for power-
on/power-off of equipment, set to defaults: ‘NO’ and ‘NO’
respectively. Refer to operator’s manual for more details.
TEST MULTIPLE RESISTANCES:
TEST GROUND WIRE RESISTANCE:
TEST POINT-TO-PT RESISTANCE:
Configures ground resistance test in MANUAL MODE. For
measurement of power cord resistance, set to defaults: ‘NO’,
‘YES’, and ‘NO’ respectively. Refer to operator’s manual for
more details.
Press More Settings to view time/date and other settings.
SET CLK:
Allows setting the internal real time clock. Press Set Clk to set
the current time and date for the device.
FORMAT:
Select date format for the internal real-time clock. Press Format
to toggle between:
• USA (MMM DD YYYY) (default)
• Europe (DD MMM YYYY)
• ISO (YYYY MMM DD)
EDIT ID:
ES601-CMMS I.D. is set at time of manufacture to "MY ES601”,
and can be edited using this utility which allows you to assign a
unique I.D. to identify ES601-CMMS within your equipment
control system. Pressing Edit ID will activate the QWERTY
touch keypad of ES601-CMMS, allowing the user to type in an
I.D. Press Enter on the keypad to save the unique name and
return to the Systems Settings Menu.
NETWORK SETUP:
Allows you to input Ethernet settings (IP address, netmask, and
gateway) for ES601-CMMS. Addresses are entered with
decimal delimiters. A touch keypad that provides for numerical
input only is displayed in response to pressing any of the
Change keys.
LABEL PRINTER SETUP:
Enables and configures output to PM Tag printer (Dymo
LabelWriter model SE300). Enter title for PM Tags and
equipment PM interval via this menu. Press Back to save
changes to PM Tag printer setup.
Quick Start Guide