Datex-Ohmeda S/5 monitors
Document No. 8001013-1
22
4.1.2
Temperature
Trouble
Cause
Treatment
Message
'TEMPERATURE ERROR'
Faulty calibration.
Perform calibration. If it does not
help, check that front panel
connectors are properly connected to
STP board.
No temperature displayed
Wrong type of probe.
Temperature out of measurable range.
Temperature calibration not protected.
Use correct probe.
The range is between 10 and 45 °C.
Set the Protection ON in the Service
Menu.