3.2
Service check
These instructions include complete procedures for a service check. The service check is
recommended to be performed after any service repair. However, the service check procedures can
also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (
Appendix A
) which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the check form should be signed after performing
the procedure.
The procedures are designed for monitors with S/5 monitor software of revision 01. However, most
of the procedures also apply to monitors, which contain some other monitor software
type/revision.
3.2.1
Recommended tools
Tool
Order No.
Notes
Pressure manometer
Adult cuff & hose
Infant cuff & hose
Screwdriver
3.2.2
Recommended parts
Part
Order No.
Notes
NIBP pump filter
57142
•
Detach the module box by removing the two screws from the back of the module. Be careful
with loose latch and spring pin for locking.
1.
Check internal parts:
−
screws are tightened properly
−
cables are connected properly
−
all IC’s that are on sockets are attached properly
−
EMC covers are attached properly
−
the calibration protection switch on the NIBP board is intact
−
tubes are not pinched and there are no sharp bends on them
Summary of Contents for S/5 M-NIBP
Page 2: ......
Page 36: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 32 This page intentionally left blank ...
Page 44: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 40 ...
Page 45: ...APPENDIX A NIBP Module M NIBP Document No 8001015 1 41 APPENDIX A ...
Page 46: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 42 ...