S/5 Airway modules
Document No. 800 1005-1
3
1.2.7
Des
Measurement range
0 to 18 %
Extended range
18 to 30 % (unspecified)
Resolution
two decimals when the AA concentration below 1.0 %
If AA concentration is below 0.10 %, 0.00 % is displayed.
1.2.8
Agent identification
Identified agents
HAL, ENF, ISO, SEV, DES
Identification time
30 seconds (typical value with pure agents)
Identification threshold
0.15 vol% (typical)
Mixture warning when minor component concentration > 0.3 vol% and > 15 %of total agent
concentration
1.2.9
Patient Spirometry
Values are valid when:
Respiratory rate
adult 4...30
pedi 4...50 breaths/min
I:E ratio
1:3 - 1:0.5
Inner diameter of ET tube is
≥
5.5 mm (adult) or 3 to 6 mm (pediatric).
1.2.10
Airway Pressure (Paw)
Accuracy
±
1.5 cmH
2
O
Resolution
1 cmH
2
O
Measuring range
-20 to +80 cmH
2
O
1.2.11
Tidal Volume (TV)
Accuracy
±
6 % or 30 ml (adult);
±
6 % or 4 ml (ped)
Resolution
1 ml
Measurement range
150 to 2000 ml (adult)
15 to 300 ml (ped)
1.2.12
Minute Volume (MV)
Resolution
0.1 l/min
Measurement range
2 to 15 l/min (adult)
0.5 to 5 l/min (ped)
1.2.13
Airway flow
Measurement range
1.5 to 100 l/min for both directions (adult)
0.25 to 25 l/min for both directions (ped)
Summary of Contents for S/5 G-AO
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1...
Page 83: ......