![Datex-Ohmeda S/5 Compact Anesthesia Monitor Technical Reference Manual Download Page 45](http://html.mh-extra.com/html/datex-ohmeda/s-5-compact-anesthesia-monitor/s-5-compact-anesthesia-monitor_technical-reference-manual_513839045.webp)
General Service Guide
4.5.3 Connection to parallel printers
A parallel printer is connected to the serial interface port connector X11 on the frame unit via the
Serial-to-Parallel Converter. The converter gets power from the connected devices.
Contact your authorized Datex-Ohmeda distributor for advice on suitable parallel printers.
NOTE: The Serial-to-Parallel Converter, order code 78030 should not be connected to the AS/3 or
CS/3 family monitors. The voltage output on the AS/3 and CS/3 monitors’ serial port would
damage the converter.
Installing the Serial-to-Parallel Converter
Table 7 Parts needed to interface a parallel printer
No Part
Order
No.
1
Common RS-232 serial modem cable
78031
2
Serial to parallel converter
78030
3
DB25 female to female gender changer
78032
4
PC to parallel printer cable
713701
1. Make sure that power is switched off on both devices.
2. Connect a standard PC-to-parallel-printer cable to the parallel printer.
3. Check the DIP-switch settings on the converter :
Position
Flow Control:
hardware
LED: Enabled
Parity/D, Bits/S, Bit:
None/8/1
D.Rate (kbps):
115,2
1
OFF
2
ON
3
OFF
4
OFF
5
OFF
6
OFF
7
ON
8
OFF
4. Connect a DB25 female-to-female gender changer between the printer cable and converter.
5. Connect a common RS-232 modem cable’s DB9 connector to the serial port connector X11
on the monitor, and the DB25 connector to the converter.
DB9F plugs
into CM1
Gender changer
plugs into
converter
DB25M end
plugs into gender
changer
Centronics
end plugs
into printer
Standard Async Modem Cable
Serial to Parallel
Converter
DB25F/DB25F
Gender Changer
Standard Parallel Printer
Adapter Cable
Parallel printer
cam
_
p1_i
nst_ser
ial
_pa
rall
e
l_
conv.vsd
1
2
3
4
Figure 13 Connection to parallel printer
Document No. M1031508
35
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...