Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
32
3.2
Service check
These instructions include complete procedures for a service check. The service check is
recommended to be performed after any service repair. However, the service check procedures can
also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (
Appendix A
) which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the check form should be signed after performing
the procedure.
The procedures are designed for monitors with S/5 monitor software of revision 01. However, most
of the procedures also apply to monitors, which contain some other monitor software
type/revision.
3.2.1
Recommended tools
Tool
Order No.
Notes
Screwdrivers
Tools for blocking internal tubes
A glass of water
Flowmeter
Multimeter
Gas Interface Cable 2.5 m
884299
Sampling line 3.0 m
73319
Spirometry tube
884101
D-lite
733950
Calibration gas
755582
3.2.2
Recommended parts
Part
Order No.
Notes
Special tube
733383
Special tube
733382
OM ref. filter
86901
Fan filter
871558
Cable tie
64001
D-fend O-ring (2 pcs)
65312
D-fend (black)
876446
Sampling line 3.0 m
73319
Extra silicon tubing
Spare constriction cassettes
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......