Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
16
2.2.2
ACX-200 measuring unit
The ACX photometer is of dual path type. The infrared light beam passes through a measuring
chamber containing the gas to be analyzed, and a reference chamber, which is free of CO
2
, N
2
O,
and AA. The measurement is made by determining the ratio between the two light intensities.
Figure 9
ACX photometer (ACX-200 measuring unit)
A filter wheel is used to control the light from an incandescent lamp that passes through the
photometer. The filters are arranged so that the light is passed sequentially:
•
first at the CO
2
absorption wavelength through the reference chamber
•
then through the measuring chamber
•
finally it is blocked completely
The same sequence is repeated at the N
2
O and anaesthetic agent gas absorption wavelengths.
After passing through the filters the light is reflected and focused by a mirror onto the infrared
detector. This detector measures the three light levels for each gas described above.
There is an optical sensor incorporated in the photometer which detects light from a reflective
surface on the filter wheel once every revolution. The pulses from this sensor are used to
synchronize the electronics to the signal from the infrared detector. A stabilizing diode measures
the temperature, which is needed to compensate for thermal drifts. The infrared detector, the
optical sensor and the stabilizing diode are mounted on the preamplifier board.
Summary of Contents for S/5 B-GAS
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Page 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Page 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Page 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Page 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Page 83: ......