Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
4
1.3
Technical specification
1.3.1
CO
2
Measurement rise time
<360 ms (from 10 to 90 %)
Gain stability
≤
0.2 %CO
2
/24 h (0 to 8 %)
≤
0.4 %CO
2
/24 h (8 to 10 %)
Gain temperature drift
≤
0.2 %CO
2
/10 °C (0 to 8 %)
≤
0.4 %CO
2
/10 °C (8 to 10 %)
Nonlinearity error
≤
0.2 %CO
2
(0 to 8 %)
≤
0.4 %CO
2
(8 to 10 %)
1.3.2
O
2
Measurement rise time
<480 ms (from 10 to 90 %)
Gain drift
≤
2 % O
2
/24 h
Gain temperature drift
≤
3 % O
2
/10 °C
Nonlinearity error
≤
2 % O
2
1.3.3
N
2
O
Measurement rise time
<360 ms (from 10 to 90 %)
Gain drift
≤
2 % N
2
O/24 h
Gain temperature drift
≤
3 % N
2
O/10 °C
Nonlinearity error
≤
2 % N
2
O
1.3.4
AA
Measurement rise time
<520 ms (from 10 to 90 %)
Gain drift
≤
0.4 % AA/24 h
Gain temperature drift
≤
0.4 % AA/10 °C
Nonlinearity error
≤
0.2 % AA
Protection against electrical shock