General specifications
Document No. M1051339
55
Transducer sensitivity:
5
µ
V/V/mmHg, 5 Vdc, max.20 mA
Pressure filter:
22 Hz (-3 dB) for P1, Art
9 Hz (-3dB) for P2, CVP, PA, ICP
10.1.7 CO
2
(option N-LCM)
Principle of operation:
Mainstream, non-dispersive infrared, single beam, single frequency, fully
chopped, ratiometric
Measurement range:
0-99 mmHg / 0-13 vol-% / 0-13 kPa
Measurement rise time:
100 ms
Accuracy
**)
0-40 mmHg /0-5.3 vol-%: ± 2 mmHg / ±0.3 vol-%
41-76 mmHg /5.3-10%: ± 5% of reading
77-99 mmHg /10-13%: ± 10% of reading
Respiration rate measurement
range:
4-150 breaths/minute
Pressure compensation:
automatic, continuous
Gas compensations:
User selectable O
2
, N
2
O or OFF
Calibration:
No routine calibration required; automatic self-calibration every 30 ms
Alarms:
adjustable high and low alarm limits for Et/Fi CO
2
and respiration rate
CO
2
sensor weight:
< 18,5 g (excluding cable)
Cable length:
3 meters
Warm-up time:
< 80 seconds from 25°C
Response time:
100 ms
**) typical value in conditions 22.5°C, 760 mmHg, CO
2
in balance air.
10.2 General
Monitor (WxDxH):
325 x 160 x 210 mm / 12.8 x 6.3 x 8.3”
Battery module (WxDxH):
290 x150 x 40 mm / 11.4 x 5.9 x 1.6”
Weight w/o batteries:
4.2 kg/9.3 lb.
Weight w/ batteries:
6.1 kg/13.5 lb.
Input voltage:
12.5 V DC
Power consumption
17.5 ... 37 W depending on the options used
Safety standards:
Designed to meet IEC 60601-1, CAN/CSA C22.2 No. 601.1-M90 and UL
2601-1, CE-marking according to Directive 93/42/EEC
Output specifications:
RS-232 computer serial output
Defibrillation synchronization signal (According to ANSI/AAMI EC13-1992.
Time interval from R peak to sync pulse output < 35 ms.)
Nurse Call
Analog output, 2 fixed and 2 freely configurable outputs
Printing:
Local printing with laser printer (PCL5), or via network.
Recorder option:
Built-in three channel thermal array recorder.
Summary of Contents for F-LM1
Page 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Page 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Page 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Page 78: ......
Page 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Page 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Page 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Page 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Page 94: ......
Page 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Page 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Page 114: ......
Page 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Page 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Page 136: ......
Page 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Page 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Page 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Page 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Page 206: ......
Page 208: ......
Page 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 224: ......
Page 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 232: ......
Page 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Page 244: ......
Page 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Page 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Page 256: ......
Page 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Page 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Page 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Page 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Page 286: ......
Page 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Page 304: ......
Page 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Page 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Page 325: ...Spare parts Document No M1051358 19 ...
Page 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Page 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Page 342: ......
Page 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Page 352: ......
Page 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Page 360: ......
Page 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Page 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Page 368: ......