![Datex-Ohmeda F-LM1 Technical Reference Manual Download Page 47](http://html.mh-extra.com/html/datex-ohmeda/f-lm1/f-lm1_technical-reference-manual_513835047.webp)
Service check
Document No. M1051339
39
6.2.7 Temperature measurement
26. Enter
the
ESTP : STP
service menu:
Modules -
ESTP : STP
Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not
increasing faster than by 50 per second. Check also that the STP board memories have
passed the internal memory test, i.e. the “RAM”, “ROM” and “EEPROM” show all OK.
27.
Check that the “Cable” and “Probe” show OFF for channel T1 when no probe is connected.
Connect a temperature test plug into the connector T1. Check that the “Cable” and “Probe”
for T1 show ON and the corresponding temperature value appears on the monitor screen.
28.
Check the temperature calibrations using temperature test plugs. If the deviation on a
temperature reading on the screen is more than 0.1 °C, calibrate the temperature channels
according to the instructions in the Technical Reference Manual.
29.
Activate the temperature test by selecting
Temp Test
from the menu and pressing the
ComWheel twice. When the message “Performing temp test” disappears from the digit field,
check that no error messages appear and “Temp error” shows OFF for both channels in the
service menu.
30.
Check that the module configuration has been set correctly. The configuration in use is
shown beside the text “Configuration” in the service menu and it can be either STP, ST or TP.
Change the configuration in the
Calibrations
menu, if necessary.
6.2.8 Invasive blood pressure measurement
31.
Check that the “Cable” and “Probe” for P1 show OFF. Plug a cable with an invasive blood
pressure transducer into the front panel connector P1 and check that the “Cable” and
“Probe” show ON and the corresponding pressure waveform appears on the screen.
Perform the same check also for the InvBP channel P2.
32.
Calibrate the InvBP channels P1 and P2 according to the instructions in the Technical
Reference Manual.
Summary of Contents for F-LM1
Page 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Page 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Page 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Page 78: ......
Page 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Page 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Page 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Page 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Page 94: ......
Page 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Page 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Page 114: ......
Page 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Page 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Page 136: ......
Page 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Page 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Page 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Page 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Page 206: ......
Page 208: ......
Page 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 224: ......
Page 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 232: ......
Page 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Page 244: ......
Page 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Page 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Page 256: ......
Page 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Page 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Page 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Page 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Page 286: ......
Page 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Page 304: ......
Page 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Page 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Page 325: ...Spare parts Document No M1051358 19 ...
Page 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Page 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Page 342: ......
Page 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Page 352: ......
Page 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Page 360: ......
Page 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Page 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Page 368: ......