![Datex-Ohmeda F-LM1 Technical Reference Manual Download Page 33](http://html.mh-extra.com/html/datex-ohmeda/f-lm1/f-lm1_technical-reference-manual_513835033.webp)
Functional check
Document No. M1051339
25
5.1.4 ECG board tests
14. Enter
the
ESTP : ECG
service menu:
Press the
Menu
button and select:
Monitor Setup - Install/Service
(Password 16-4-34)
-
Service View
(Password 26-23-8)
- Modules - ESTP : ECG
Check from the Service Data that:
−
the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not increasing
faster than by 50 per second.
−
the ECG/RESP board memories have passed the internal memory test, i.e. “RAM”,
“ROM” and “EEPROM” all state
OK
.
15.
Check that the module mains power frequency “Power Freq” value has been set according
to the supply frequency. If necessary, change the setting by selecting
Power Freq
.
16.
Check that the “Resp Available” and “Resp measurement” both show
ON
.
NOTE: The Resp measurement shows
OFF,
if
Resp
is not selected to the screen setup. To
setup the screen, press the
Menu
key and select:
Monitor Setup - Waveform Fields
or
Digit Fields.
"
5.1.5 STP board tests
17. Enter
the
ESTP : STP
service menu:
−
Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not
increasing faster than by 50 per second.
−
Check that the STP board memories have passed the internal memory test, i.e. “RAM”,
“ROM” and “EEPROM” all show OK.
18.
Check that the protection for temperature calibration is on:
“Protect key” text in the menu should state
OFF
.
“Protect mode” text should state
ON
.
If necessary, change the protection mode in the
Calibrations
menu.
"
Summary of Contents for F-LM1
Page 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Page 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Page 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Page 78: ......
Page 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Page 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Page 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Page 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Page 94: ......
Page 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Page 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Page 114: ......
Page 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Page 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Page 136: ......
Page 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Page 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Page 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Page 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Page 206: ......
Page 208: ......
Page 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 224: ......
Page 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 232: ......
Page 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Page 244: ......
Page 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Page 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Page 256: ......
Page 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Page 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Page 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Page 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Page 286: ......
Page 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Page 304: ......
Page 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Page 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Page 325: ...Spare parts Document No M1051358 19 ...
Page 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Page 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Page 342: ......
Page 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Page 352: ......
Page 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Page 360: ......
Page 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Page 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Page 368: ......