Datex-Ohmeda S/5 Light Monitor
2.3 Adjustments and calibrations
2.3.1 CO
2
calibration check
Welch Allyn recommends that the CO
2
sensor should be checked at least once annually to ensure
its accuracy.
Equipment needed:
•
CO
2
Calibration Check Adapter Kit (P/N 902305)
•
Calibration Gas 5.00 % CO
2
(P/N 755580)
•
Patient Airway Adapter.
755534
Obstruct
hole with
the tube
755533
When using Datex-Ohmeda calibration gas regulator note the following:
•
With regulator P/N 755533 remove the tube with Y-piece. Connect the silicone tube (comes
with the adapter) on the nozzle of the regulator.
•
With regulator P/N 755534 slide the silicone tube until it blocks also the little hole on the
nozzle. This will ensure that the CO2 concentration rises high enough in the calibration
check adapter.
To check the mainstream CO
2
sensor
1.
Turn the monitor ON.
2.
Change the mode of operation to MainSt.
3.
Change the CO
2
units of measure to % CO
2
.
4.
Fix the mainstream sensor to the patient airway adapter.
5.
Let the CO
2
sensor warm up for 15 minutes.
6.
At the end of the 15-minute warm-up period, observe the reading of
CO
2
in %. The CO
2
reading should be between 0.0 % - 0.3 % with well
ventilated room air.
7.
Fix the mainstream sensor to the CO
2
calibration airway adapter.
8.
Feed 5.00 % CO
2
gas (e.g. Datex-Ohmeda Quick Cal Calibration Gas, order No. 755580) to
the free end of the silicon tube (included in the kit P/N 902305). Introduce a steady stream
of gas until FiCO
2
and EtCO
2
readings are same and observe the reading.
The CO
2
reading should be 5 %
±
0.3 % for a 5 % gas.
9.
If the values do not fall within the acceptable ranges, contact your local GE Healthcare
distributor for further information.
Document No. 8000060-1
8
Summary of Contents for F-LM1
Page 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Page 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Page 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Page 78: ......
Page 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Page 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Page 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Page 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Page 94: ......
Page 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Page 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Page 114: ......
Page 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Page 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Page 136: ......
Page 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Page 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Page 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Page 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Page 206: ......
Page 208: ......
Page 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 224: ......
Page 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 232: ......
Page 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Page 244: ......
Page 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Page 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Page 256: ......
Page 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Page 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Page 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Page 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Page 286: ......
Page 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Page 304: ......
Page 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Page 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Page 325: ...Spare parts Document No M1051358 19 ...
Page 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Page 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Page 342: ......
Page 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Page 352: ......
Page 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Page 360: ......
Page 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Page 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Page 368: ......