Service menu
5.1.3 NIBP pulse valve menu
Start test
Start test
is for starting and
Stop test
is for stopping the test.
Set valve
Set Valve
lets you adjust the opening of the pulse valve.
NOTE:
Parameter values in Service Data are for reference only.
Pulse valve data
See NIBP service menu in chapter
for information on general items
Pressure, Zero, Protect
handle, Calibr. prot., +12 V power, AD0
to
AD7
as well as
Timeouts
etc.
Pulse valve checking
Wrap an adult cuff around a pipe and connect the cuff to the monitor. Select the
Start test
and
push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The
module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and
displays it. The test can be manually stopped by selecting
Stop test
.
The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First
select Set Valve and push the ComWheel. See the pulse valve value and adjust it by turning the
ComWheel. Then push the ComWheel to confirm the value.
The "
Interval 240 mmHg -> 50 mmHg
" time should be less than 60 seconds when the valve is
'150' and less than 10 when fully opened (255). When fully closed (0), the system should be
airtight and the pressure does not drop. Depending on an individual, the pulse valve may remain
closed up to approx. value 100.
If the measured time deviates much from those above, then the pulse valve or its tubes are faulty.
Document No. 8000060-1
55
Summary of Contents for F-LM1
Page 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Page 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Page 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Page 78: ......
Page 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Page 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Page 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Page 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Page 94: ......
Page 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Page 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Page 114: ......
Page 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Page 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Page 136: ......
Page 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Page 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Page 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Page 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Page 206: ......
Page 208: ......
Page 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 224: ......
Page 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 232: ......
Page 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Page 244: ......
Page 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Page 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Page 256: ......
Page 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Page 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Page 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Page 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Page 286: ......
Page 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Page 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Page 304: ......
Page 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Page 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Page 325: ...Spare parts Document No M1051358 19 ...
Page 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Page 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Page 342: ......
Page 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Page 352: ......
Page 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Page 360: ......
Page 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Page 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Page 368: ......