S/5 Anesthesia Monitor and S/5 Critical Care Monitor
30
Document No. 8001004-1
4.
The Command board PCB can be detached by disconnecting the cable connectors and
removing the screws (3).
Rev. 00, 01, 02
1.
The backlight unit as well as the LCD Display component can be detached by removing the
screws (4) at their corners.
2.
The Command board PCB can be detached by disconnecting the cable connectors and
removing the screws (2).
3.
The Power supply board is attached to the rear panel by four (4) screws.
Reassembly of the LCD Display is made in reversed order. Make sure that all connectors are
connected properly and cables are not pinched between covers.
Replacing the Backlight Tubes, D-LCC10A/W
1.
Disconnect the folio cables (2 pcs) by opening connector’s front part in cable direction.
2.
Disconnect the other cables from the LCD interface board.
3.
Unscrew 5 screws and draw out the lid.
4.
Detach the connector of the backlight tube from the inverter board.
5.
Unfasten those two screws that holds the backlight tube and draw out it with the clamp.
6.
Locate the new backlight tube, fasten the screws, and connect it to the invert board.
7.
Attach all the parts, and make sure that the folio cables are connected properly. Make sure
that the folio cables are at the bottom and straight before locking the connector.
Adapter board (892424) jumpers on D-LCC10A/W
The right position for jumper X3 is 2-3.
Position for jumper X13 according to cable lenght:
1-2
0.5...2.5 m
2-3
> 2.5...10.0 m
NOTE: If you change the cable to a long (10 m) or to a short (2.5 m or shorter) one, you may have to
change jumper setting inside the display cabin in order to get a clear picture. Let an authorised
person to do this.
3.3.2
LCD Display, D-LCC15
Replacing the Backlight Tubes
1.
Disconnect the display cables from the Central Unit and lift off the display together with the
display stand.
2.
Remove the hole covers and 6 screws from the back corners and bottom line of the display
case. Separate the front panel from the back panel of the monitor .
3.
Remove 8 screws attaching the display case to he front display case. Take display case out
of the front cover and place it face down on a nonabrasive clean surface.
Summary of Contents for D-VHC17
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv ...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1 ...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G ...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1 ...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1 ...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1 ...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1 ...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1 ...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1 ...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1 ...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1 ...