3.2.4
Video Display, D-VSC21
Tools needed
Command Board, Central Unit and parameter modules
•
Turn the monitor to STBY.
1.
Check that the display cover and the picture tube screen are intact.
?
2.
Check that all display screws are tightened properly.
?
3.
Check that the display power cord is intact.
?
4.
Disconnect the display video cable from the Display controller board and check that the
connector pins are clean and straight. Check that the thumb screws on the video cable
connector are intact.
Reconnect the video cable and lock it properly.
?
•
Make sure that the display has been powerless at least two minutes, then turn the monitor
on.
5.
Check that the picture on the display screen is positioned correctly. Readjust the picture
position, if necessary.
?
6.
Check that the brightness and contrast are adjusted correctly. Readjust them, if necessary.
?
7.
Check that all the colors on the screen are clear.
NOTE: Internal degaussing (demagnetisation) for clearing the colors takes place
automatically when the monitor is turned on.
If necessary, perform extra degaussing using an external degaussing coil.
?
8.
Perform electrical safety check and leakage current test.
?
Summary of Contents for D-VHC17
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv ...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1 ...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G ...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1 ...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1 ...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1 ...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1 ...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1 ...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1 ...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1 ...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1 ...