S/5 Anesthesia Monitor and S/5 Critical Care Monitor
Document No. 8002948
12
3.3.2
Performing Factory Reset
NOTE: The Factory Reset is necessary after downloading of monitor software and after replacing the
CPU board or SRAM/Timekeeper battery.
NOTE: The Factory Reset will restore all your customized defaults, including language selection, to
factory defaults.
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and password (16-4-34).
3.
Select
Service
and password (26-23-8).
4.
Select
Set/Test
and perform
Factory Reset
.
5.
The monitor will perform an automatic restart. After the restart is completed, restart the
monitor also manually by the On/Standby –switch.
3.4
Adjustments and calibrations
No calibrations or adjustments are needed on the CPU board, B-CPU5/B-CPU4.