![Datex-Ohmeda AS/3 Compact Monitor Technical Reference Manual Download Page 31](http://html.mh-extra.com/html/datex-ohmeda/as-3-compact-monitor/as-3-compact-monitor_technical-reference-manual_513828031.webp)
General Service Guide
23
Document No. 896 623
3.4
AS/3 Memory Board, B-CMMEM Installation (w/ B-CPU2/3)
1.
Make sure that the power is turned off on the Compact Monitor and unplug the Power Cord.
Press and hold the Service Reset button on the rear panel for at least five seconds until the
service reset indicator LED is switched off.
2.
Unscrew the six screws holding the back plate in position and remove the plate.
WARNING
The Compact Monitor is always energized by the internal battery. Do not touch
any exposured wiring or conductive surface with metallic object while the back
plate is off or electric failure may result.
CAUTION
The Compact Monitor is always energized by the internal battery. A short circuit may
cause internal damage.