Datex-Ohmeda AS/3 Compact Monitor, CS/3 Compact Monitor
20
Document No. 896 623
3.3.8
Replacing the Software Cartridge of B-CPU2/3
NOTE:
All user settings will be lost after replacing the software cartridge.
1.
Make sure that the power is turned off on the Compact Monitor and unplug the power cord.
Press and hold the service reset button on the rear panel for at least five seconds until the
service reset indicator LED is switched off.
2.
Unscrew the two screws holding the software cartridge in position and remove the cartridge.
3.
Insert the new software cartridge into the empty slot and firmly press the cartridge in
position. Secure the cartridge (two screws).
4.
Reconnect the power cord.
5.
Perform factory reset.
NOTE:
After changing the software cartridge the start-up time is considerably longer when the
monitor is started for the first time.