Troubleshooting A-ELEC
Document No. 8502856
D-45
Discharge/recharge cycling test
Run the ADU on batteries until it turns itself off and then let it recharge. Observe that the discharge
current for each pack is about 1A and that the charge current is 450 to 500 mA. Repeat this cycle if
necessary. This will level out the differences in the batteries. This test is recommended when new
batteries are installed.
NOTE: When replacing batteries in ADU all 4 must be replaced at the same time. Batteries supplied
by Datex-Ohmeda should always be used in ADU, because we are testing and selecting the batteries
to each others. ADU problems due to locally purchased batteries we do not take any responsibility
for.
Unexpected reset
In a reset condition, the display shows the message ´System failure - Unexpected Reset, please call
for service´.
If an Unexpected reset occurs the Electronic Ventilator and the Fresh Gas Control units continue to
deliver gas as set, but you will be unable to adjust the ventilator settings. Fresh gas flows and
anesthetic agent settings can be adjusted, but the settings are not displayed on the screen.
How the reset system of the ADU works
When an ADU is started the following happens: As soon as power is applied the watchdog (located at
the Digipower board) activates the reset signal to the CPU. When the +5V line rises above 4,65V an
internal timer in the watchdog keeps the reset active for 200 ms to ensure that the CPU will be
properly reset after power up. When the main SW starts to execute it sends a start command to the A-
EV1. If main SW gets a good response from the A-EV1 it continues with the start-up of the ADU.
However if it receives a message from the A-EV1 saying that it is already running, this is interpreted by
the main SW as an error. Main SW should start the A-EV1, if it is already running this is an indication
that the CPU has done a warm start. This will result in the error messages on the screen and prevent
further operations of the ADU. During normal program execution main SW must do a watchdog kick at
least once every second to prevent it from doing a reset of the CPU.
Display description if a reset occurs
When a reset condition occurs, the ADU displays the following:
1.
The screen blanks out.
2.
After approximately 10 seconds, the “Datex-Ohmeda” is shown.
3.
After another 20 seconds, the screen will appear as if in a normal start-up.
4.
After another 20 seconds, the message “System Failure - Unexpected Reset, Please call for
service” appears in red.
Service recommended actions if a reset occurs
•
Check service error log
•
Check Voltages on diagnostic pages
•
Run HW watchdog test from service page if test fails replace.
−
A. Replace the B-CPU board and repeat the test.
−
B. Replace the Digipower board and repeat the test.
•
On the B-CPU2 board check that the SW cassette is correctly installed and that the insulation is
present.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...