5 Troubleshooting
1006-0453-000 05/04
5-27
5.5.2 No display troubleshooting
Start
Turn on System
Check Contrast
Adjustment
Service Mode
No
Yes
No
Check Display
Cable to CPU
Connections
Yes
Replace EL
Display Panel
Is display dim
but information
can be seen?
Problem
continues?
Yes
Is there a
continuous or
cycling audio
tone?
Yes
No
Is AC LED
power light on?
Check system
circuit breaker
Check AC
power cable
No
Disconnect J5 connector from CPU.
Perform continuity check on
System Switch.
Yes
Does the unit have
Integrated CPU?
Disconnect 6 pin cable J200
from CPU to Power Supply.
Verify 12.5V and 16V power present.
Yes
Replace
Power Supply
Voltages
present?
No
Yes
Disconnect 50 pin cable J2
from CPU to Power Supply.
Verify all voltages present.
No
Replace
Power Supply
Board
Replace CPU Board
No
Yes
Done
Was there single
audio tone on
power-up?
Yes
No
Check/Replace harness.
Replace System Switch.
No
Problem
continues?
Done
No
Yes
Problem
continues?
Replace CPU Board
Replace CPU Board
Replace CPU Board
AB.29.095
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......