7 User Maintenance
WARNING
To help prevent fires:
•
Only use lubricants approved for anesthesia or O2
equipment, such as Krytox.
•
Do not use lubricants that contain oil or grease. They
may burn or explode in high O2 concentrations.
•
All materials used to cover the system must be made
from antistatic (conductive) materials. Static electricity
can cause fires.
•
Desiccated (dehydrated) absorbent material may
produce dangerous chemical reactions when exposed
to inhalation anesthetics. Adequate precautions
should be taken to ensure that absorbent does not dry
out. Turn off all gases when finished using the system.
w
Obey infection control and safety procedures. Used
equipment may contain blood and body fluids.
w
Moveable parts and removable components may present
a pinch or a crush hazard. Use care when moving or
replacing system parts and components.
In this section
Repair policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Maintenance summary and schedule . . . . . . . . . . . . . . . . . . 7-2
Breathing system maintenance. . . . . . . . . . . . . . . . . . . . . . . 7-3
O2 cell replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
O2 cell calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Zeroing flow sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Prevent water buildup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...