Pneumatic problems
System
Problem
Solution
High-pressure leak test
fails.
Controls are not set correctly.
Make sure that no gas is
flowing, turn off the auxiliary
flowmeter, and repeat the test.
Incorrect cylinder connection.
Make sure that there is only
one cylinder gasket, the gasket
is in good condition, and the
connection is tight.
Low-pressure leak test
fails with a vaporizer on.
The vaporizer is not correctly
installed.
Correctly install the vaporizer.
The vaporizer filler is loose
(funnel fill type vaporizer).
Tighten the filler.
Vaporizer port o-rings (external)
are damaged or not installed.
Install new o-rings.
A vaporizer malfunction (the leak
stops if a different vaporizer is
used in the same manifold
position).
Send the vaporizer to an
approved service center for
repair.
A port valve malfunction (the
leak continues if a different
vaporizer in the same manifold
position).
Contact a Datex-Ohmeda
trained service representative
to repair the vaporizer
manifold.
Low-pressure leak with a
vaporizer off.
Anesthesia machine problem.
Contact a Datex-Ohmeda
trained service representative.