Electromagnetic compatibility (EMC)
WARNING
Changes or modifications to this equipment not expressly
approved by the manufacturer could cause EMC issues
with this or other equipment. Contact the manufacturer for
assistance. This device is designed and tested to comply
with applicable regulations regarding EMC as follows.
w
Use of portable phones or other radio frequency (RF)
emitting equipment (that exceed electromagnetic
interference levels specified in IEC 60601-1-2) near the
system may cause unexpected or adverse operation.
Monitor operation when RF emitters are in the vicinity.
w
Use of other electrical equipment on or near this system
may cause interference. Verify normal operation of
equipment in the system before use on patients.
Guidance and
manufacturer’s
declaration -
electromagnetic
emissions
The system is suitable for use in the specified electromagnetic
environment. The customer and/or the user of the system should
assure that is used in an electromagnetic environment as described
below.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The system uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The system is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...