Aespire View
10-22
M1132382
PCV-VG
1. Paw waveform
2. Tinsp
3. Texp
4. Variable pressure to deliver desired TV
5. PEEP
6. Flow waveform
7. TV
Figure 10-8 • PCV-VG waveforms
In PCV-VG, a tidal volume is set and the ventilator delivers that
volume using a decelerating flow and a constant pressure. The
ventilator will adjust the inspiratory pressure needed to deliver the set
tidal volume breath-by-breath so that the lowest pressure is used.
The pressure range that the ventilator will use is between the PEEP +
2 cmH2O level on the low end and 5 cmH2O below Pmax on the high
end. The inspiratory pressure change between breaths is a maximum
of ±3 cmH2O.
This mode will deliver breaths with the efficiency of pressure
controlled ventilation, yet still compensate for changes in the patient’s
lung characteristics. PCV-VG begins by first delivering a volume
breath at the set tidal volume. The patient’s compliance is determined
from this volume breath and the inspiratory pressure level is then
established for the next PCV-VG breath.
PCV-VG settings include:
•
Pinsp - sets the target airway pressure.
•
RR - sets the rate of mechanically driven breaths.
•
I:E - sets the amount of inspiration to expiration ratio.
•
PEEP - sets the positive end expiratory pressure.
•
Pmax - sets the maximum airway pressure.
•
Rise Rate - sets the amount of time to attain Pinsp.
2
1
3
4
5
6
7
AB
.9
8
.03
4
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...