Aespire View
8-14
M1132382
High-pressure leak
test
Note
Datex-Ohmeda recommends testing one cylinder at a time.
1. Turn on the system.
2. Disconnect pipeline supplies.
3. Turn off the auxiliary O2 flowmeter and the venturi suction (if
equipped).
4. Open the cylinder.
5. Record the cylinder pressure.
6. Close the cylinder.
•
If the cylinder pressure decreases more than 690 kPa
(100 psi) in one minute there is a significant leak.
7. To repair a leak, install a new cylinder gasket and tighten the tee
handle.
8. Repeat the leak test. If the leak continues, do not use the system.
How to attach equipment to the top of the machine
WARNING
Do not exceed the top shelf weight limit of 34 kg (75 lb).
Check the stability of the system in its final configuration.
Make sure that the weight is evenly distributed throughout
the system.
1. Locate the clips or slots.
2. Install the straps.
3. Fully tighten the straps.
4. Make sure the straps hold the equipment in position.
WARNING
Fully tighten the straps. If straps are not fully tightened,
equipment can fall off the top of the machine.
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...