3.1 Post-service checkout
After servicing the 7100 ventilator, perform the following service mode Calibrations:
•
O
2
Calibration
•
Zero Flow and Airway Sensors
•
Adjust Drive Gas Regulator
•
Airway Sensor Span
•
PEEP Valve Calibration
•
Inspiratory Valve Calibration
•
Pressure Sensitivity
Then, you must complete the checkout procedure for the entire machine:
•
the 7100 ventilator (refer to Section 3.1.1)
•
the anesthesia machine (refer to Section 3.1.2)
•
and all the accessories and options.
w
WARNING
You must perform all post-service checks after maintenance or service of
the ventilator. Failure to do so may result in patient injury.
w
WARNING
All components and accessories must be connected correctly. All hoses
and cables must be properly connected before returning the anesthesia
machine to clinical use. Failure to do so may result in patient injury.
3.1.1 Test the 7100
ventilator
Perform the
Preoperative Checkout Procedure
that applies:
•
in Part 2 of the Aestiva/5 7100 User’s Reference Manual.
•
in Part 2 of the S/5 Aespire User’s Reference Manual.
3.1.2 Test the anesthesia
machine
The 7100 ventilator is an integral component of the Aestiva or the Aespire anesthesia
machine.
To be certain the ventilator is functioning correctly, test the entire system.
For the proper checkout procedures, refer to the following document that applies:
•
the
Aestiva Anesthesia Machine Technical Reference Manual
.
•
the
Aespire Anesthesia Machine Technical Reference Manual
.