![Datex-Ohmeda 7100 Technical Reference Manual Download Page 119](http://html.mh-extra.com/html/datex-ohmeda/7100/7100_technical-reference-manual_513823119.webp)
7 Repair Procedures
1006-0836-000 02/03
7-13
7.3.3 Pneumatic engine
components
(Aespire machine)
Regulator (A)
Also inspect the two o-rings that seal it to the manifold. Replace as necessary.
PEEP Safety Valve (B)
Ensure the valve you are installing (white dot) is in this location.
PEEP Valve (C)
Ensure the valve you are installing (two blue dots) is in this location.
Supply Pressure Switch (D)
Also inspect the o-ring that seal it to the manifold. Replace as necessary.
Transfer harness wire to top and middle tabs; bottom tab open.
Inspiratory Flow Valve (E)
Note orientation of the flow valve. Also inspect the two o-rings that seal it to the
manifold. Replace as necessary.
Inlet Filter (F)
Install the filter (with the course side DOWN) into the recess in the manifold.
Place the o-ring into its groove.
F
A
B
D
E
C