7100 Anesthesia Ventilator
7-4
02/03 1006-0836-000
7.2 Control Module
The control module is mounted on either a folding mount or on a repositionable
arm. In addition to the power cord, signals between the control module and the
machine mounted components are sent through three cables that pass
through the side of the machine (and through the arm if equipped) and
connect to the rear of the control module.
To replace the fuses
and the fan filter
•
To replace the fuses (
A
), first remove the power cord retainer (
B
) and
the power cord (
C
).
•
To replace the filter (
D
), remove the filter retainer (
E
). If the Control
Module is mounted on an arm, remove the control panel from the arm
to access the filter.
To remove the control module
1. Disconnect the three interface cables (
F
) and the power cord (C
C
C
C).
2. Remove the screws that attach the control module to the mounting
brackets.
3. Remove the control module (C
C
C
CM
M
M
M).
4. Reassemble in reverse order.
5. Refer to Section 7.1 for installing software and updating the control board
configuration.