7100 Anesthesia Ventilator
6-4
02/03 1006-0836-000
6.3 MOPV pressure relief valve test
w
WARNING
Objects in the breathing system can stop gas flow to the patient. This
can cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
• Make sure that there are no test plugs or other objects caught in the
breathing system.
6.3.1 Test setup
1. Disassemble the breathing system to the point where you can remove the
exhalation valve (do not remove the exhalation valve on Aespire machine).
2. Plug the inspiratory flow (drive pressure) port with a stopper.
Figure 6-2 • Occluding the inspiratory flow port
6.3.2 Test procedure
1. Set the system switch to On and enter the Service Mode.
2. Select “Diagnostic Tests/Tools”; select “Valves - Test Tool”.
3. Select “Set Inspiratory Valve” and activate a flow of 10 L/min.
4. Carefully listen for the MOPV relief weight to be relieving and “popping off”
from its seat (a purring sound). This indicates the valve is functioning
correctly.
5. Set the system switch to Standby.
6. Remove the stopper from the inspiratory flow port.
7. Reassemble the system.
8. Perform the Preoperative Checkout Procedure
(refer to the respective User’s Reference manual).
Plug Inspiratory Flow (Drive Pressure) Port
Aestiva anesthesia machine
Aespire anesthesia machine