![Dataq DI-2108 User Manual Download Page 29](http://html1.mh-extra.com/html/dataq/di-2108/di-2108_user-manual_3691315029.webp)
DI-2108 Hardware Manual
Controls, Indicators, and Connections
23
4-20mA Current Loop Measurements
Use the following diagram for 4-20mA current loop measurement connections.
The shunt resistor should be placed on the low side of the circuit as shown below:
Status
D0
D1
D2
D3
D4
D5
D6
Model DI-2108
USB
Button
+5V GnD
CH8
CH7
CH6
CH5
CH4
CH3
CH2
CH1
±100V Max
0-25V Max
Count
Rate
Record
Event
Drive
www.dataq.com
±10V F.S.
Shunt Resistor
Model R250
+
-
4-20 mA
current source
Shunt Resistor
I
4-20mA
Current Loop
Transmitter
Loop Power
Supply
To DI-2108
Analog Input Channel
+
-
+
-