84
H-Class
Advanced Entry Calibration (continued)
Step
Action
Displayed Message
Comment
F
Use the buttons to set the
Gain Number to 00 and
then press ENTER.
Record the sensor reading
as a Label Value for Gain
Number 00 in a table (32
rows by four columns,
with headings similar to
those shown below.)
TRAN SENSOR GAIN
yyy (0 - 31)
00
This is the Label Value for
a gain setting of 00,
where “yyy” represents
the current sensor
reading.
Sample Calibration Table
Gain Number
Label Value
TOF Value
Difference Value
00 252
01
02
…
31
Step
Action
Displayed Message
Comment
G
Use the buttons to
increment the Gain
Number by one and then
press ENTER. Record the
Label Value. Repeat this
process for each Gain
Number.
TRAN SENSOR GAIN
yyy (0 - 31)
01
These are Label Values,
where “yyy” represents
the current sensor
reading.
Sample Calibration Table
Gain Number
Label Value
TOF Value
Difference Value
00 252
01 250
02 248
… …
31 09