Datacolor CONDITIONER™ (MB2) 06/2017
page 75/126
When operating the cabinet without water connection, turn off humidity control in the “set-
points” menu (chap. 6.3) in order to avoid humidity alarms.
11.2 State of alarm
1.
Visual indications in Normal display: alarm message, screen header flashing in red color
2.
Audible alert, if the buzzer is enabled (chap. 11.5).
Normal display in state of alarm (example for
cabinet with lights).
(a)
Screen header flashing in red color and
showing the alarm message
(b) Alarm
icon on the bottom of the screen:
change to the list of active alarms and
alarm acknowledgement
(c)
If applicable, information icon in the screen
header. Indication of a certain condition
11.3 Resetting an alarm, list of active alarms
Normal display in state of alarm (example for
cabinet with lights).
Press the
Alarm
icon
List of active alarms.
Press the
Reset alarm
icon.
Pressing the
Reset alarm
icon mutes the buzzer for all active alarms. The icon then disappears.
•
Acknowledging while the alarm condition persists: Only the buzzer turns off. The visual alarm indica-
tion remains on the controller display. The alarm remains in the list of active alarms.
When the alarm condition has ended, the visual alarm indication is automatically cleared. The alarm is
then no longer in the list of active alarms.
•
Acknowledging after the alarm condition has ended: The buzzer and the visual alarm indication are
reset together. The alarm is then no longer in the list of active alarms.
•
The zero-voltage relay alarm output resets together with the alarm.
(a)
(c)
(b)