Shimadzu LC-10/20 Pump
4 Using the control module
4.4 Shimadzu LC-10/20 Pump Setup
Shimadzu LC-10/20 Pump Setup
dialog (accessible through the
System
Configuration
dialog) is used to set the connection to the pump, select it's
type and set some other basic parameters.
Fig 10: Shimadzu LC-10/20 Pump Setup
Port
Sets the COM port to which the
Shimadzu LC- 10/20 Pump
pump is
connected.
Pump Model
Allows to select from several types of Shimadzu pumps.
LC-10ADvp
and
LC-10ATvp
models are supported at the moment.
Note:
If any
LC- 20
type pump should be connected, it is necessary to set the
LC-10 emulation mode
on the pump (for more details see the chapter
Installation Procedure
on pg
3
)..
Valve Model
Allows to select the valve model that may be installed on the pump to
create low pressure gradient.
Sys Mode
Sets the pump operation mode.
Isocratic
and
Low Pres. Grad.
options are
available.
Note:
The pump hardware must be set accordingly (for more details see the
chapter
Installation Procedure
on pg
3
)..
Solvents
Sets the number of solvents when the
Low Pres. Grad.
mode is used.
Max Pressure
Sets the maximum pressure allowance of the pump. This value is sent to
the pump and when the pressure exceeds this value, pump will react on it.
The pressure check is also present in
Clarity
, in the
Method Setup - LC
Gradient - Gradient Options
dialog. Value of
Max Pressure
set there
- 13 -