4 Using the control module
Clarity Controls
4.2.8 Method Setup - GC - Other
Fig 16: Method Setup - GC - Other
This tab allows to set the behavior of some parts of the
Shimadzu GC-
201x
chromatograph.
Shutdown
Defines the behavior of the chromatograph after the analysis end.
Keep State
The chromatograph will do nothing when the analysis is completed.
Set Off after analysis
The chromatograph performs the
System Off
function
(check
Shimadzu GC-201x
manual for more details). This function is useful
for switching the chromatograph off after the last sample in the
sequence. When the
Clarity Instrument
is set to send a method
without prompting in the
System Configuration - Instrument Method
Sending
dialog and there is another method opened in the
Instrument
window at the analysis end, the chromatograph will be turned off by
the current method and then turned on by the sending of the
instrument method.
Caution:
Be careful when switching off instruments without automatic pressure
control and automatic flow control - the manually controlled gases flow is
not switched off, which may end in the leakage of possibly dangerous
gases (hydrogen) into the laboratory environment.
- 20 -