Signal Polarity
This field is only present when using the dual detector in the
Dual
mode or
one of TCD detectors. Sets the polarity of the dual data, or in other words,
the method of data calculating (either the signal of the right cell minus the
signal of the left cell (polarity
-
) or vice versa (polarity
+
)). Available for
TCD and WDFID detector only.
Range
Applies on the detector signal before its processing to the analog output in
the
Shimadzu GC-201x
. This value serves for the compensation of the
overly high signals.
Current
Serves for entering the value of current set on the detector. Possible range
and units depend on the type of the detector -
FTD
detector has the range
of values
0.00-10.00 pA
,
ECD
detector
0.00-2.00 nA
and
TCD
detector
0-
200 mA
. This field is only present for Electron capture (ECD), Thermal
conductivity (TCD) and Flame thermoionic (FTD and WFTD) detectors.
Signal Output Port
Specifies the signal output of the particular detector.
Shimadzu GC-201x
provides up to 4 signal outputs set in the
Shimadzu GC- 201x Setup -
Signals
dialog
Power Controller
This field is only present on tabs of Flame thermoionic detectors (FTD and
WFTD).
- 13 -