4 Using the control module
Clarity Controls
20
LAN it will be necessary to set the
Baud Rate
lower. Default
value is
115200
.
The baud rate set in this dialog must correspond to the baud
rate value set in the chromatograph.
Name
Specifies the name of the chromatograph.
Autodetect
Reads the hardware configuration from the
Shimadzu GC-
2014C
chromatograph and sets it to the
Shimadzu GC-2014C
Setup
dialog.
Piping tab
Fig. 17 Shimadzu GC-2014C Setup - Piping
Car 1 (..3)
Allows to set the possible operation modes for carrier gasses.
For
Shimadzu GC-2014C
chromatographs (without
automatic flow control function), only the
SPLITTER
and
PACKED
options will be used, others are saved for
instruments equipped with AFC control.
Inj 1 (2)
Lists the type of the injector(s) installed.
Det 1 (..4)
Lists all detectors and other signal sources (e.g. temperature
of the measuring cell) from which the signals will be
transferred into
Clarity
. Each
Det X
field with any type of
detector other than
None
will have its own sub-tab with