4 Using the control module
15
turned off by the current method and then turned
on by the sending of the instrument method.
Be careful when switching off instruments without automatic
pressure control (
APC
-
GC-2014C
) and automatic flow control
(
AFC
- all
GC-2014
chromatographs) - the manually
controlled gasses flow is not switched off, which may end in
the leakage of possibly dangerous gasses (hydrogen) into the
laboratory environment.
Zero at ready
When checked, the chromatograph performs the autozero
action after it becomes
READY
.
Auto flame ON
When checked, the chromatograph will try to ignite the
flame in the detector after sufficient temperature and gasses
pressure was reached.
Re-ignite
While checked, the chromatograph will try to re-ignite the
flame in the detector when it is quenched. This may be
important during the start of the chromatograph as the auto
ignition may not be successful on the first try.