Shimadzu GC-2014C
4 Using the control module
11
Interference Filter
Possible values are
P
,
S
and
Sn
. This field is only present
when using the Flame photometric detector (FPD or WFPD).
4.2.4 Time Events
Menu Path:
Instrument -> Method -> GC Control
Fig. 10 Method Setup - GC - Time Events
This tab allows to set the events that should happen during
the run based on the analysis time. This includes events like
switching the detector polarity, range or current (where
applicable), setting temperatures to
detector/injector/auxiliary zones etc. To delete a row, simply
delete the value in the
Time
column on that row.
The table should be filled with the
Time
of the event in the
first column (in minutes, decimal format - 2 minutes 45
seconds should be set as 2.75), the event type in the
Event
column, event value in the
Value
column and the
Line
selected as set on the
Shimadzu GC-2014C Setup - Line Config
tab. Possible events are:
Det 1(..4) Polarity
Sets the polarity of the detector (if the polarity for the given
detector has any sense). Polarity
+
corresponds to the value
1
, polarity
-
corresponds to the value
2
.
Det 1(..4) Range