Shimadzu GC-2014C
4 Using the control module
9
GC-2014C
hardware manual for more details on these
injection modes).
Sampling Time [min.]
Sets the sampling time in minutes for the
Splitless
Sampling
Mode
.
4.2.3 DET 1 (..4)
Menu Path:
Instrument -> Method -> GC Control
Fig. 9 Method Setup - GC - DET 1
This tab allows to set the parameters for a detector
configured in the
Shimadzu GC-2014C Setup
dialog. Particular
fields may vary, according to the type of the detector set.
On/Off
Switches the detector on and off. The detector switched off
gives
0
as a resulting signal.
Filter Time Const.
The value set here influences the processing of the detector
signal. It works as a noise filter, the higher the constant, the
lower the noise. On the other hand, usage of the higher
values will flatten peaks and may cause lower or smaller
peaks to disappear.