4 Using the control module
Clarity Controls
8
Time
. To delete a row, simply delete the value in the
Heat
Rate
column on that row.
Temperature Zones
Sets the temperature and maximum temperature of the
zones that have been set in the
Shimadzu GC-2014C Setup -
Heater Ports
dialog (for more details see chapter
Shimadzu
GC-2014C Setup
on pg.
19
).
Description
Description of the method (this description is not sent to or
read from GC, it serves for the information purposes only).
4.2.2 INJ 1(2)
Menu Path:
Instrument -> Method -> GC Control
Fig. 8 Method Setup - GC - INJ 1
This tab governs the settings concerning the injection port of
the
Shimadzu GC-2014C
chromatograph. Up to two
injection ports may be configured and used at the same
time. The tab is only present when the
SPL
operation is
selected in the
Shimadzu GC-2014 Setup - Piping
dialog (for more
details see chapter
Piping tab
on pg.
20
) for a given injector
port.
Split Mode
Defines the operation mode for the injector port. The
possible options are
Split
,
Splitless
and
Direct
(see
Shimadzu